A proof-of-concept 3D-printed hydrogel aortic valve displayed a low transvalvular pressure gradient of 14–24 mmHg and endured a cyclic pressure of ~80 mmHg for at least 180,000 cycles.
A novel 3D-printed hydrogel bioprosthetic valve demonstrates favorable mechanical properties, hemocompatibility, and durability in vitro, offering a potential future alternative for heart valve replacement.
Bioprosthetic cardiac valves are efficient replacements to clinically treat the valvular heart diseases including stenosis and regurgitation, while these often suffer from the issues of mechanical mismatch, biocompatibility, thrombus, degradation, etc. Here, we report a big step toward the biomechanically compatible hydrogel heart valves by the reversible addition–fragmentation chain transfer polymerization compatible three-dimensional (RAFT-3D) printing tough hydrogels. The biocompatible hydrogel valves have favorable mechanical properties matching with native valves and long-term stability and durability owing to the wide and robust hydrogen-bonding networks, on which the heparin-like polymer with −SO3– is readily grafted through RAFT polymerization, endowing outstanding hemocompatibility including low hemolysis, anticoagulation, and acceptable inflammatory response. The proof-of-concept hydrogel aortic valve displayed a low transvalvular pressure gradient of 14–24 mmHg, which efficiently regulates the flow direction and regurgitation and, more importantly, endures a cyclic pressure of ∼80 mmHg for at least 1.8 × 105 cycles yet maintaining a stable transvalvular pressure gradient without any damage to the configuration and performance. It is believed that the engineered hydrogel cardiac valves pave a significant way toward addressing the complications of cardiovascular diseases.
Wu et al. (Tue,) conducted a other in Valvular heart diseases. RAFT-3D printed hydrogel bioprosthetic valves was evaluated on Transvalvular pressure gradient and durability. A proof-of-concept 3D-printed hydrogel aortic valve displayed a low transvalvular pressure gradient of 14–24 mmHg and endured a cyclic pressure of ~80 mmHg for at least 180,000 cycles.