Key result
The implanted telemetric pulmonary artery pressure monitor accurately measured pressures with a bias of -0.43 compared to reference catheters and safely tracked disease progression over 30 days.
Why the study?
Pulmonary hypertension causes progressive right heart failure, and proactive pulmonary artery pressure monitoring reduces hospitalization in heart failure, motivating evaluation of a novel implanted monitor in PH models.
Does an implanted pulmonary artery pressure monitor accurately detect clinically relevant changes in pressure in porcine models of pulmonary hypertension?
Does an implanted pulmonary artery pressure monitor accurately detect clinically relevant changes in pressure in porcine models of pulmonary hypertension?
Mean Difference: -0.43 (95% CI -5.3–4.4)
A novel implanted pulmonary artery pressure monitor accurately and safely detected clinically meaningful hemodynamic changes in porcine models of acute and chronic pulmonary hypertension.
Implanted PA pressure monitoring may detect changes in PH; hypothesis-generating and requires prospective outcome trials before any clinical adoption.
Aims Pulmonary hypertension (PH) is associated with significant morbidity and mortality and leads to progressive right heart failure. In patients with PAH, haemodynamic parameters measured at catheterisation relate to clinical worsening events, in patients with heart failure proactive pulmonary artery pressure based therapeutic intervention reduces hospitalisation. We therefore investigated use of a novel implanted pulmonary artery (PA) pressure monitor to detect clinically relevant changes in pressure in large animal models of pulmonary hypertension (PH). Methods and Results Prototype pulmonary artery pressure sensors (Endotronix) were implanted using standard interventional techniques. Acute PH was induced by infusion of thromboxane A2 in domestic swine. Over a physiological range pressure monitors remained concordant to reference catheter (bias −0.43, 95%CI-5.3-4.4). Chronic PH was induced by i.p. injection of monocrotaline. Implanted pressure sensors demonstrated a gradual rise in PA pressure over 30 days (baseline: 20.7+/-0.4 vrs day-30: 31.74+/-1.4, p<0.01). Pressure sensor derived readings matched reference catheter at baseline and day-30. Pressure sensors remained stable and no adverse events were identified by clinical and histological examination. Conclusions The development of PA pressure monitors provide long-term haemodynamic data that identified clinically meaningful changes in pulmonary artery pressure. In addition to proactive heart failure management, such devices may be used to optimise or personalize patient therapy, investigate aspects of physiology and pathology essential to the understanding of disease and provide the opportunity to assess therapeutic interventions in clinical studies.
No takes yet. Share an insight, caveat, or question.
Rothman et al. (2020) studied Pulmonary hypertension (n=13). Telemetric pulmonary artery pressure monitoring system (Endotronix) vs. Reference catheter was evaluated on Concordance of pressure monitor to reference catheter (bias) (bias -0.43, 95% CI -5.3-4.4). The implanted telemetric pulmonary artery pressure monitor accurately measured pressures with a bias of -0.43 compared to reference catheters and safely tracked disease progression over 30 days.