Key result
HbE and HbD traits cause clinically significant measurement interference in up to ~22% of HbA1c assays.
Answers are grounded in this paper and cite their sources.
BACKGROUND: Glycohemoglobin (GHB), reported as hemoglobin (Hb) A(1c), is a marker of long-term glycemic control in patients with diabetes and is directly related to risk for diabetic complications. HbE and HbD are the second and fourth most common Hb variants worldwide. We investigated the accuracy of HbA(1c) measurement in the presence of HbE and/or HbD traits. METHODS: We evaluated 23 HbA(1c) methods; 9 were immunoassay methods, 10 were ion-exchange HPLC methods, and 4 were capillary electrophoresis, affinity chromatography, or enzymatic methods. An overall test of coincidence of 2 least-squares linear regression lines was performed to determine whether the presence of HbE or HbD traits caused a statistically significant difference from HbAA results relative to the boronate affinity HPLC comparative method. Deming regression analysis was performed to determine whether the presence of these traits produced a clinically significant effect on HbA(1c) results with the use of +/-10% relative bias at 6% and 9% HbA(1c) as evaluation limits. RESULTS: Statistically significant differences were found in more than half of the methods tested. Only 22% and 13% showed clinically significant interference for HbE and HbD traits, respectively. CONCLUSIONS: Some current HbA(1c) methods show clinically significant interferences with samples containing HbE or HbD traits. To avoid reporting of inaccurate results, ion-exchange chromatograms must be carefully examined to identify possible interference from these Hb variants. For some methods, manufacturers' instructions do not provide adequate information for making correct decisions about reporting results.
No takes yet. Share an insight, caveat, or question.
Little et al. (2008) studied Diabetes (HbA1c testing). Presence of HbE or HbD traits vs. HbAA results was evaluated on Clinically significant effect on HbA1c results (+/-10% relative bias at 6% and 9% HbA1c). The presence of HbE and HbD traits caused clinically significant measurement interference in 22% and 13% of 23 evaluated HbA1c methods, respectively.
Synapse has enriched 10 related papers. Consider them for comparative context: