Key result
Elevated fetal hemoglobin levels >20% artificially lowered HbA1c results from the Primus CLC 330/385, Siemens DCA2000, and Tosoh 2.2+ assays, but not the Bio-Rad Variant II and Tosoh G7.
Why the study?
Does elevated fetal hemoglobin interfere with HbA1c results measured by common assay methods compared to the IFCC reference method?
Observational
Does elevated fetal hemoglobin interfere with HbA1c results measured by common assay methods compared to the IFCC reference method?
Answers are grounded in this paper and cite their sources.
Elevated fetal hemoglobin (>20%) artificially lowers HbA1c results in certain boronate-affinity and other common assay methods, highlighting the need for physician awareness of potential interference.
Hemoglobin A1c (HbA1c) is an important indicator of risk for complications in patients with diabetes mellitus. Elevated fetal hemoglobin (HbF) levels have been reported to interfere with results of some HbA1c methods, but it has generally been assumed that HbA1c results from boronate-affinity methods are not affected by elevated HbF levels. None of the previous studies used the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method as the comparative HbA1c method. We, therefore, measured HbA1c in samples with normal and elevated HbF levels by several common assay methods and compared the results with those of the IFCC reference method.HbF levels of more than 20% artificially lowered HbA1c results from the Primus CLC 330/385 (Primus Diagnostics, Kansas City, MO), Siemens DCA2000 (Siemens Healthcare Diagnostics, Tarrytown, NY), and Tosoh 2.2+ (Tosoh Bioscience, South San Francisco, CA), but not the Bio-Rad Variant II (Bio-Rad Laboratories, Hercules, CA) and Tosoh G7. Physicians and laboratory professionals need to be aware of potential interference from elevated HbF levels that could affect HbA1c results, including those from boronate-affinity methods.
No takes yet. Share an insight, caveat, or question.
Rohlfing et al. (2008) conducted an observational in Diabetes mellitus. Elevated fetal hemoglobin (HbF) levels vs. Normal HbF levels was evaluated on HbA1c results compared to the IFCC reference method. Elevated fetal hemoglobin levels >20% artificially lowered HbA1c results from the Primus CLC 330/385, Siemens DCA2000, and Tosoh 2.2+ assays, but not the Bio-Rad Variant II and Tosoh G7.
Synapse has enriched 10 related papers. Consider them for comparative context: