Key result
A risk-based reflex protocol for creatinine testing saved approximately $40,000 per year with 0 clinically significant false-negative and 5 false-positive results out of 104,530 measurements.
Why the study?
Does a risk-based reflex testing protocol for creatinine measurement reduce costs while maintaining patient safety in outpatients?
Observational (n=104,530)
Does a risk-based reflex testing protocol for creatinine measurement reduce costs while maintaining patient safety in outpatients?
A risk-based reflex protocol for creatinine testing (Jaffe method reflexed to enzymatic for eGFR 55-65) safely reduces laboratory costs.
Answers are grounded in this paper and cite their sources.
Risk-based reflex creatinine testing may cut lab costs safely in outpatients; leaves open broader adoption pending prospective validation.
BACKGROUND: The Jaffe and enzymatic methods are the 2 most common methods for creatinine measurement. The Jaffe method is less expensive but subject to interferences. Some laboratory scientists have called for the Jaffe method to be retired. OBJECTIVE: To determine the most cost-effective and safe protocol for creatinine measurement. METHOD: We performed a retrospective database review of all outpatient creatinine measurements for 1 year, testing the risk-based reflex testing protocol we had implemented for creatinine measurement. Samples were first measured using the Jaffe method and were reflexed to the enzymatic method if the estimated glomerular filtration rate (eGFR) was between 55 and 65 mL per min per 1.73 m2. RESULTS: There were 104,530 creatinine measurements, of which 11,420 (10.9%) were reflexed to the enzymatic method. The Jaffe method had a positive bias of 0.08 mg per dL (-6.14 mL/min/1.73 m2 eGFR). A total of 3.4% of the paired reflexed specimens were discordant. Also, 133 (1.2%) of the Jaffe results were classified as false negatives and 3411 (29.9%) were classified as false positives. None of the false-negative results and 5 of the false-positive results were considered clinically significant. Using the reflex protocol saved approximately $40,000 per year. CONCLUSIONS: The reflex protocol for creatinine measurement can reduce costs with acceptable risk.
No takes yet. Share an insight, caveat, or question.
Lehman et al. (2018) conducted an observational in Outpatient creatinine testing (n=104,530). Risk-based reflex testing protocol (Jaffe reflexed to enzymatic) was evaluated on Cost-effectiveness and safety (discordance, false positives/negatives, cost savings). A risk-based reflex protocol for creatinine testing saved approximately $40,000 per year with 0 clinically significant false-negative and 5 false-positive results out of 104,530 measurements.
Synapse has enriched 10 related papers. Consider them for comparative context: