Key result
Magnetic resonance imaging in patients with an implantable loop recorder caused no harm or device malfunction, but artifacts mimicking arrhythmias were seen in all 4 patients.
Why the study?
Does magnetic resonance imaging (MRI) cause harm or device malfunction in patients with an implantable loop recorder (ILR)?
Case Report (n=4)
Does magnetic resonance imaging (MRI) cause harm or device malfunction in patients with an implantable loop recorder (ILR)?
Answers are grounded in this paper and cite their sources.
MRI scanning of patients with implantable loop recorders is safe and does not affect device functionality, but frequently produces artifacts that mimic tachyarrhythmias.
BACKGROUND: The implantable loop recorder (ILR) is a useful tool in the diagnosis of syncope. Our understanding of their functional and safety profile in interfering environments such as magnetic resonance imaging (MRI) becomes increasingly important as they become more prevalent. METHODS: We report four patients with an ILR who underwent MRI. The ILR memory was cleared before MRI and no changes were made to programmed settings. Device interrogation took place immediately after the scan. Patients were surveyed for device movement and heating, in addition to cardiopulmonary symptoms after their MRI. RESULTS: Following MRI scanning, all patients were asymptomatic and no device movement or heating was observed. In addition, the functionality of the device remained unaffected. Artifacts mimicking arrhythmias were seen in all ILR patients regardless of the type of MRI scan. CONCLUSIONS: MRI scanning of ILR patients can be performed without harm to patient or device, but artifacts that could be mistaken for a tachyarrhythmia are seen frequently.
No takes yet. Share an insight, caveat, or question.
Wong et al. (2008) conducted a case report in Implantable loop recorder (n=4). Magnetic resonance imaging (MRI) was evaluated on Device movement, heating, cardiopulmonary symptoms, and device functionality. Magnetic resonance imaging in patients with an implantable loop recorder caused no harm or device malfunction, but artifacts mimicking arrhythmias were seen in all 4 patients.
Synapse has enriched 10 related papers. Consider them for comparative context: