Key result
Magnetocardiography detected significantly lower left atrial pseudo-current increase at peak exercise in PAF patients compared to healthy volunteers and marathon runners (0.4 vs 0.8 vs 1.1 pT, p<0.001).
Why the study?
The role of magnetocardiography in detecting left atrial dysfunction in patients with paroxysmal atrial fibrillation is unknown.
Does magnetocardiography detect left atrial dysfunction in patients with paroxysmal atrial fibrillation compared to healthy subjects?
Observational (n=70)
Does magnetocardiography detect left atrial dysfunction in patients with paroxysmal atrial fibrillation compared to healthy subjects?
Magnetocardiography can non-invasively detect left atrial dysfunction during exercise, effectively differentiating patients with paroxysmal atrial fibrillation from healthy subjects.
Answers are grounded in this paper and cite their sources.
Lower LA pseudo-current on exercise MCG signals PAF-related dysfunction; hypothesis-generating and should not yet change practice.
BACKGROUND: Magnetocardiography (MCG) is a non-invasive technique and to characterize the magnetic field, a pseudo-current conversion was used. The role of MCG in detecting left atrial (LA) dysfunction in patients with paroxysmal atrial fibrillation (PAF) is unknown. OBJECTIVE: The aim of this study was to evaluate LA function using MCG in patients with PAF and healthy subjects, to identify possible indices to diagnose PAF. METHODS: We enrolled a total of 70 subjects including 26 healthy volunteers (group 1) and 22 marathon runners (group 2) who did not exhibit any cardiac abnormalities, and 22 patients with PAF (group 3) which was documented by electrocardiography (ECG). Spatiotemporal activation graph (STAG) in base-apex and left-right direction was reconstructed. The maximum value of LA pseudo-current under rest and peak exercise were measured between the end of the P wave and beginning of the Q wave. RESULTS: LA pseudo-current increase at peak exercise in PAF patients was significantly lower than in healthy volunteers and marathon runners (0.4±0.3 pT in group 3 vs. 0.8±0.3 pT in group 1 vs. 1.1±0.5 pT in group 2, p < 0.001). PAF patients had less pseudo-current increase in STAG at peak exercise than healthy volunteers and marathon runners (46% of 26 PAF patients, 81% of 22 healthy subjects vs. 81% of 22 marathon runners, p = 0.002). Sensitivity, specificity, and the area under the receiver-operator characteristics curve of LA pseudo-current increase at peak exercise for differentiating PAF patients from healthy subjects were 77%, 92%, and 0.896. CONCLUSIONS: MCG can provide important non-invasive information for detecting LA dysfunction in PAF patients. Therefore, MCG may help in differentiating PAF patients from healthy subjects.
No takes yet. Share an insight, caveat, or question.
Her et al. (2019) conducted an observational in Paroxysmal atrial fibrillation (n=70). Magnetocardiography (MCG) vs. Healthy volunteers and marathon runners was evaluated on Left atrial pseudo-current increase at peak exercise (p=< 0.001). Magnetocardiography detected significantly lower left atrial pseudo-current increase at peak exercise in PAF patients compared to healthy volunteers and marathon runners (0.4 vs 0.8 vs 1.1 pT, p<0.001).