Key result
Regional upstroke tracking algorithms reduced mean errors for local carotid pulse wave velocity estimation from 12.23% to 7.13% compared with time fiduciary point methods.
Why the study?
In ultrasonic transit time-based local PWV estimation, time fiduciary point locations in pulse wave upstrokes are inconsistent due to reflected waves and ultrasonic noise.
Does regional upstroke tracking improve the accuracy and reproducibility of ultrasound-based local pulse wave velocity estimation in carotid arteries compared to time fiduciary point methods?
Does regional upstroke tracking improve the accuracy and reproducibility of ultrasound-based local pulse wave velocity estimation in carotid arteries compared to time fiduciary point methods?
Regional upstroke tracking algorithms significantly improve the accuracy and reproducibility of ultrasound-based local carotid pulse wave velocity estimation compared to traditional time fiduciary point methods.
Answers are grounded in this paper and cite their sources.
May refine carotid PWV estimates in vascular studies; leaves open clinical translation pending validation.
Pulse wave velocity (PWV) is the most important index for quantifying the elasticity of an artery. The accurate estimation of the local PWV is of great relevance to the early diagnosis and effective prevention of arterial stiffness. In ultrasonic transit time-based local PWV estimation, the locations of time fiduciary point (TFP) in the upstrokes of the propagating pulse waves (PWs) are inconsistent because of the reflected waves and ultrasonic noise. In this study, a regional upstroke tracking (RUT) approach that involved identifying the most similar TFP-centered region in the upstrokes is proposed to detect the time delay for improving the local PWV estimation. Five RUT algorithms with different tracking points are assessed via simulation and clinical experiments. To quantitatively evaluate the RUT algorithms, the normalized root-mean-squared errors and standard deviations of the estimated PWVs are calculated using an ultrasound simulation model. The reproducibility of the five RUT algorithms based on 30 human subjects is also evaluated using the Bland-Altman analysis and coefficient of variation (CV). The obtained results show that the RUT algorithms with only three tracking points provide greater accuracy, precision, and reproducibility for the local PWV estimation than the TFP methods. Compared with the TFP methods, the RUT algorithms reduce the mean errors from 12.23% ± 3.10% to 7.13% ± 2.31%, as well as the CVs from 21.76% to 13.39%. In conclusion, the proposed RUT algorithms are superior to the TFP methods for local carotid PWV estimation.
No takes yet. Share an insight, caveat, or question.
Deng et al. (2019) studied Arterial stiffness (n=30). Regional upstroke tracking (RUT) algorithms vs. Time fiduciary point (TFP) methods was evaluated on Mean errors for local PWV estimation. Regional upstroke tracking algorithms reduced mean errors for local carotid pulse wave velocity estimation from 12.23% to 7.13% compared with time fiduciary point methods.
Synapse has enriched 10 related papers. Consider them for comparative context: