Key result
A 2-step granular echocardiographic approach combining color and continuous-wave Doppler parameters achieved excellent inter-observer agreement (86% of cases, κ = 0.79) for grading paravalvular leak severity after transcatheter aortic valve implantation.
Why the study?
Does a granular echocardiographic approach improve the reproducibility of assessing paravalvular regurgitation after TAVI?
Observational (n=50)
Blinded to clinical and procedural data
Does a granular echocardiographic approach improve the reproducibility of assessing paravalvular regurgitation after TAVI?
A granular algorithm combining color and continuous wave Doppler parameters yields excellent reproducibility for assessing paravalvular leak after TAVI by transthoracic echocardiography.
Paravalvular leak (PVL) after transcatheter aortic valve implantation (TAVI) is challenging to quantitate. Transthoracic echocardiography (TTE) is the main tool used for the assessment of PVL but is modestly reproducible. We sought to develop a reproducible echocardiographic approach to assess PVL in the post-TAVI setting. Four observers independently analyzed eleven parameters of PVL severity in 50 pre-discharge TTE studies performed after TAVI. The parameters included color-Doppler parameters [jet circumferential extent (CE) and planimetered vena contracta area in the short-axis view and jet breadth and qualitative features in the long-axis views], continuous-wave Doppler parameters [jet velocity time integral (VTI) and pressure half time (PHT)], quantitative Doppler parameters (regurgitation volume and fraction and effective regurgitant orifice area), aortic diastolic flow reversal and valve stent eccentricity. Intraclass correlation coefficient (ICC) and coefficient of variation (CV) for numerical parameters and kappa coefficient (κ) for categorical parameters were calculated for inter- and intra-observer comparisons. Inter-observer ICC was highest and CV lowest for CE (0.88 and 0.36), jet origin breadth (0.82 and 0.39), jet qualitative features in long-axis views (0.87 and 0.26), jet VTI (0.87 and 0.04) and PHT (0.73 and 0.10). Similar results were found in intra-observer comparisons. A 2-step granular approach combining the most reproducible parameters was used to grade PVL by the four observers. Inter-observer agreement was achieved in 86 % of cases (κ = 0.79). Combining color Doppler and continuous wave Doppler parameters in a granular algorithm yields excellent reproducibility of PVL assessment by TTE.
No takes yet. Share an insight, caveat, or question.
Abdelghani et al. (2016) conducted an observational in Paravalvular leak after transcatheter aortic valve implantation (TAVI) (n=50). 2-step granular echocardiographic approach was evaluated on Inter-observer agreement on paravalvular leak grade (κ = 0.79, p=<0.001). A 2-step granular echocardiographic approach combining color and continuous-wave Doppler parameters achieved excellent inter-observer agreement (86% of cases, κ = 0.79) for grading paravalvular leak severity after transcatheter aortic valve implantation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: