Key result
Low-voltage impedance drops ~15 ohms over 60 minutes post-S-ICD implant, reliably approximating high-voltage impedance.
Why the study?
Data are limited regarding the trajectory of noninvasive low voltage impedance during S-ICD insertion and its proximity to high voltage shock impedance from DFT testing.
Does noninvasive low voltage impedance at 60 minutes post-implantation accurately approximate high voltage impedance in patients undergoing S-ICD insertion?
Observational (n=53)
No
Does noninvasive low voltage impedance at 60 minutes post-implantation accurately approximate high voltage impedance in patients undergoing S-ICD insertion?
Intraprocedural low-voltage impedance predictably declines within 60 minutes of S-ICD implantation and closely approximates high-voltage impedance, potentially offering a noninvasive justification to defer standard defibrillation threshold testing.
Answers are grounded in this paper and cite their sources.
LV impedance at 60 min approximates HV values post-S-ICD; leaves open whether routine DFT testing can be safely deferred.
BACKGROUND: Following subcutaneous implantable cardioverter-defibrillator (S-ICD) insertion, induction of ventricular fibrillation (VF) and defibrillation threshold testing (DFT) is the standard of practice to assess system performance. In contrast, DFT testing is not routinely performed after transvenous ICD (T-ICD) insertion, supported by literature that emphasizes negative clinical repercussions and reliability of noninvasive predictors of successful defibrillation. There is a paucity of data for both the trajectory of noninvasive low voltage (LV) impedance measurements at S-ICD insertion through the time of potential DFT testing and for the proximity of LV impedance measurements and high voltage (HV) shock impedance from 65J DFT or 10J testing. OBJECTIVE: We aim to characterize the trend in LV impedance from initial S-ICD connection (LV T0) to final intraprocedural impedance (LV T4) and to compare LV impedance with HV impedance. METHODS: Data from patients who underwent S-ICD implantation between July 2022 and March 2024 were analyzed. LV impedance measurements were collected during implant and HV impedance measurements were collected if 65J DFT/10J testing was performed. Differences in mean impedances at each time point were evaluated using paired t-tests. For those who underwent 10J or 65J DFT testing, LV impedance was compared with HV impedance. RESULTS: The primary analytic sample included 53 patients, with a mean age of 51 ± 15.7 years; 16 (30.2%) were female. LV impedance decreased by a mean of 14.5 ± 9.8 ohms (18.3%; p <0.001) from S-ICD connection (T0) to the final intraprocedural measurement, approximately 60 min postimplant (T4). Among 35 patients who underwent HV testing (65J DFT or 10J), LV impedance at T4 and HV impedance did not differ significantly (1.7 ± 6.3 ohms; p = 0.116). CONCLUSION: Our findings demonstrate a predictable reduction in LV impedance from S-ICD connection to subsequent intraprocedural measurements. We also found that LV impedance at LV T4, the final intraprocedural measurement, did not significantly differ from contemporaneous HV impedance. Together, these findings indicate that LV impedance at implant can be expected to decline, and that 60-min postimplant LV impedance may reliably approximate HV impedance. Evaluation of circumstances in which LV impedance could influence the decision to perform DFT testing is warranted.
No takes yet. Share an insight, caveat, or question.
Kontopidis et al. (2025) conducted an observational in Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation (n=53). Intraprocedural low voltage (LV) impedance measurement during S-ICD insertion vs. Baseline (T0) LV impedance or contemporaneous high voltage (HV) impedance was evaluated on Change in low voltage (LV) impedance from S-ICD connection (T0) to 60 minutes postimplant (T4) (Mean decrease of 14.5 ohms (18.3%), p=<0.001). Intraprocedural low voltage impedance significantly decreased by a mean of 14.5 ohms over 60 minutes post-S-ICD implant, reliably approximating high voltage impedance.
Synapse has enriched 10 related papers. Consider them for comparative context: