Key result
A 10 J shock impedance in sinus rhythm correlated strongly with a 65 J defibrillation impedance during intermuscular subcutaneous defibrillator implantation.
Why the study?
Defibrillation testing is recommended during S-ICD placement, but whether a 10 J shock impedance in sinus rhythm compares with 65 J defibrillation impedance using an intermuscular technique was evaluated.
Does 10 J shock impedance in sinus rhythm correlate with 65 J defibrillation impedance during intermuscular S-ICD implantation?
Cohort (n=37)
No
Does 10 J shock impedance in sinus rhythm correlate with 65 J defibrillation impedance during intermuscular S-ICD implantation?
A 10 J shock in sinus rhythm correlates well with standard 65 J defibrillation impedance, suggesting it may be sufficient to predict defibrillation efficacy during intermuscular S-ICD implantation.
Answers are grounded in this paper and cite their sources.
Low-energy impedance may guide S-ICD positioning; leaves open whether it alters energy selection or outcomes in practice.
INTRODUCTION: Defibrillation testing (DT) is recommended during the subcutaneous defibrillator (S-ICD) placement. We sought to compare 10 J shock impedance in sinus rhythm (SR) with 65 J defibrillation impedance and evaluate device position on a postimplant chest X-ray (CXR) using an intermuscular (IM) technique. METHODS: Consecutive S-ICD implantations between 12/2019 and 12/2020 at The Ohio State University were reviewed. All implantations were performed using a two-incision IM technique. Standard DT with 65 J shock and 10 J shock in SR were performed unless contraindicated. The PRAETORIAN score was calculated based on CXR. RESULTS: ) between 10 J shock impedance and 65 J shock impedance was 0.84. The mean of an impedance difference was 1.6 ± 4.8 Ω (minimum - 11 and maximum 8). Postimplant CXR was available for 33 out of 37 patients (89.2%). The PRAETORIAN score was less than 90 in all patients and the mean score was 32.7 ± 8.8. CONCLUSION: We demonstrated that 10 J shock impedance in SR correlated well with 65 J defibrillation impedance during IM S-ICD implantation. An IM implantation technique provides excellent generator location on postimplant CXR. The IM technique combined with 10 J shock in SR may be sufficient to predict and ensure the defibrillation efficacy of the S-ICD.
No takes yet. Share an insight, caveat, or question.
Okabe et al. (2021) conducted a cohort in Subcutaneous defibrillator (S-ICD) implantation (n=37). 10 J shock impedance in sinus rhythm vs. 65 J defibrillation impedance was evaluated on Correlation between 10 J shock impedance and 65 J shock impedance (Correlation 0.84). A 10 J shock impedance in sinus rhythm correlated strongly with a 65 J defibrillation impedance during intermuscular subcutaneous defibrillator implantation.