Key result
Wearable flexible ultrasound array reliably measures local pulse wave velocity, averaging ~3.9 m/s in healthy subjects.
Why the study?
Large and cumbersome ultrasound probes require appropriate positioning and pressure maintenance, introducing usability constraints for local pulse wave velocity monitoring.
A newly developed flexible ultrasound array demonstrates feasibility and reliability for non-invasive, bandage-like monitoring of local pulse wave velocity to assess arterial stiffness.
Answers are grounded in this paper and cite their sources.
Supports feasibility of flexible ultrasound for local PWV in healthy subjects; leaves open clinical validation in larger cohorts.
Pulse wave velocity (PWV) measured at a specific artery location is called local PWV, which provides the elastic characteristics of arteries and indicates the degree of arterial stiffness. However, the large and cumbersome ultrasound probes require an appropriate sensor position and pressure maintenance, introducing usability constraints. In this paper, we developed a light (0.5 g) and thin (400 μm) flexible ultrasound array by encapsulating 1-3 composite piezoelectric transducers with a silicone elastomer. It can capture the distension waveforms of four arterial positions with a spacing of 10 mm and calculate the local PWV by multi-point fitting. This is illustrated by in vivo experiments, where the local PWV value of five normal subjects ranged from 3.07 to 4.82 m/s, in agreement with earlier studies. The beat-to-beat coefficient of variation (CV) is 12.0% ± 3.5%, showing high reliability. High reproducibility is shown by the results of two groups of independent measurements of three subjects (the error between the mean values is less than 0.3 m/s). These properties of the developed flexible ultrasound array enable the bandage-like application of local PWV monitoring to skin surfaces.
No takes yet. Share an insight, caveat, or question.
Xu et al. (2022) studied Healthy (normal blood pressure) (n=5). Flexible ultrasound array for local pulse wave velocity monitoring was evaluated on Local pulse wave velocity (PWV). The flexible ultrasound array successfully measured local pulse wave velocity in healthy subjects, yielding an average of 3.9 m/s with a beat-to-beat coefficient of variation of 12.0%.