Key result
Quantification of myocardial blood flow with 82Rb PET demonstrated excellent concordance with 15O-water PET over the entire flow range, validating its accuracy for clinical use even at high flow rates.
Why the study?
Does 82Rb PET accurately quantify myocardial blood flow and flow reserve compared to 15O-water PET in healthy controls and patients with mild CAD?
Cross-Sectional (n=33)
Single-blind
Yes
Does 82Rb PET accurately quantify myocardial blood flow and flow reserve compared to 15O-water PET in healthy controls and patients with mild CAD?
82Rb PET with a newly derived correction for nonlinear extraction accurately quantifies myocardial blood flow and flow reserve compared to the 15O-water gold standard.
Answers are grounded in this paper and cite their sources.
May support 82Rb PET for MBF quantification in mild CAD; leaves open outcome validation before routine adoption.
PURPOSE: Quantification of myocardial blood flow (MBF) with generator-produced (82)Rb is an attractive alternative for centres without an on-site cyclotron. Our aim was to validate (82)Rb-measured MBF in relation to that measured using (15)O-water, as a tracer 100% of which can be extracted from the circulation even at high flow rates, in healthy control subject and patients with mild coronary artery disease (CAD). METHODS: MBF was measured at rest and during adenosine-induced hyperaemia with (82)Rb and (15)O-water PET in 33 participants (22 control subjects, aged 30 ± 13 years; 11 CAD patients without transmural infarction, aged 60 ± 13 years). A one-tissue compartment (82)Rb model with ventricular spillover correction was used. The (82)Rb flow-dependent extraction rate was derived from (15)O-water measurements in a subset of 11 control subjects. Myocardial flow reserve (MFR) was defined as the hyperaemic/rest MBF. Pearson's correlation r, Bland-Altman 95% limits of agreement (LoA), and Lin's concordance correlation ρ (c) (measuring both precision and accuracy) were used. RESULTS: Over the entire MBF range (0.66-4.7 ml/min/g), concordance was excellent for MBF (r = 0.90, [(82)Rb-(15)O-water] mean difference ± SD = 0.04 ± 0.66 ml/min/g, LoA = -1.26 to 1.33 ml/min/g, ρ(c) = 0.88) and MFR (range 1.79-5.81, r = 0.83, mean difference = 0.14 ± 0.58, LoA = -0.99 to 1.28, ρ(c) = 0.82). Hyperaemic MBF was reduced in CAD patients compared with the subset of 11 control subjects (2.53 ± 0.74 vs. 3.62 ± 0.68 ml/min/g, p = 0.002, for (15)O-water; 2.53 ± 1.01 vs. 3.82 ± 1.21 ml/min/g, p = 0.013, for (82)Rb) and this was paralleled by a lower MFR (2.65 ± 0.62 vs. 3.79 ± 0.98, p = 0.004, for (15)O-water; 2.85 ± 0.91 vs. 3.88 ± 0.91, p = 0.012, for (82)Rb). Myocardial perfusion was homogeneous in 1,114 of 1,122 segments (99.3%) and there were no differences in MBF among the coronary artery territories (p > 0.31). CONCLUSION: Quantification of MBF with (82)Rb with a newly derived correction for the nonlinear extraction function was validated against MBF measured using (15)O-water in control subjects and patients with mild CAD, where it was found to be accurate at high flow rates. (82)Rb-derived MBF estimates seem robust for clinical research, advancing a step further towards its implementation in clinical routine.
No takes yet. Share an insight, caveat, or question.
Prior et al. (2012) conducted a cross-sectional in Coronary Artery Disease (n=33). 82Rb positron emission tomography vs. 15O-water positron emission tomography was evaluated on Concordance of myocardial blood flow (MBF) measurements between 82Rb and 15O-water (Lin's concordance correlation ρc 0.88, 95% CI 0.82-0.94). Quantification of myocardial blood flow with 82Rb PET demonstrated excellent concordance with 15O-water PET over the entire flow range, validating its accuracy for clinical use even at high flow rates.