Key result
In patients undergoing S-ICD generator replacement, the mean defibrillation threshold was 27.4 ± 14.3 J, with the majority having a threshold ≤30 J regardless of BMI.
Why the study?
Concerns were raised about high defibrillation thresholds during S-ICD generator replacements, possibly related to fibrosis, but prospective data were lacking.
Observational (n=45)
The defibrillation threshold during S-ICD generator replacement is generally low (≤30 J in most patients), easing concerns about high DFTs due to fibrosis and suggesting the device's shock output could be safely reduced.
DFT remains low at elective S-ICD replacement; supports reevaluation of default output but leaves open need for randomized confirmation.
IntroductionThe first step-down defibrillation studies in the subcutaneous implantable cardioverter-defibrillator (S-ICD) described a defibrillation threshold (DFT) of 32.5 ± 17.0 J and 36.6 ± 19.8 J. Therefore, the default shock output of the S-ICD was set at 80 J. In de novo implants, the DFT is lower in optimally positioned S-ICDs. However, a retrospective analysis raised concerns about a high DFT in S-ICD replacements, possibly related to fibrosis.ObjectiveWe aimed to find the DFT in patients undergoing S-ICD generator replacement.MethodsThis prospective study enrolled patients who underwent S-ICD generator replacement with subsequent defibrillation testing. A pre-specified defibrillation testing protocol was used to determine the DFT, defined as the lowest shock output that effectively terminated the induced ventricular arrhythmia.ResultsA total of 45 patients were enrolled, 6.0 ± 2.1 years after initial implant. Mean DFT during replacement in the total cohort was 27.4 ± 14.3 J. In patients with a body mass index (BMI) 18.5–25 kg/m2 (N = 22, BMI 22.5 ± 1.6), median DFT was 20 J (IQR 17.5–30). In 18/22 patients, the DFT was ≤30 J and 5/22 patients were successfully defibrillated at 10 J. One patient with hypertrophic cardiomyopathy had a DFT of 65 J. In patients with a BMI >25 kg/m2 (N = 23, BMI 29.5 ± 4.2), median DFT was 30 J (IQR 20–40). In 15/23 patients, the DFT was ≤30 J and 4/23 patients had a successful defibrillation test at 10 J.ConclusionsThis study eases concerns about a high DFT after S-ICD generator replacement. The majority of patients had a DFT ≤30 J, regardless of BMI, suggesting that the shock output of the S-ICD could be safely reduced.
No takes yet. Share an insight, caveat, or question.
Stuijt et al. (2023) conducted an observational in S-ICD generator replacement (n=45). S-ICD generator replacement was evaluated on Defibrillation threshold (DFT). In patients undergoing S-ICD generator replacement, the mean defibrillation threshold was 27.4 ± 14.3 J, with the majority having a threshold ≤30 J regardless of BMI.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: