Key result
The cuff-less SOMNOtouch NIBP demonstrated a mean deviation of 0.99 mmHg for systolic and 3.03 mmHg for diastolic BP compared to a cuff-based device during 24-hour ambulatory measurements.
Why the study?
Cuff-less devices can alleviate discomfort and nocturnal arousal reactions from conventional cuff-based ambulatory blood pressure monitoring, but their clinical validity for 24-h measurements in children and adolescents needed assessment.
Does the cuff-less SOMNOtouch NIBP device provide accurate 24-h blood pressure measurements compared to the cuff-based Mobil-O-Graph in children aged 5-17?
Observational (n=90)
Does the cuff-less SOMNOtouch NIBP device provide accurate 24-h blood pressure measurements compared to the cuff-based Mobil-O-Graph in children aged 5-17?
Mean Difference: 0.99
The cuff-less SOMNOtouch NIBP device provides accurate 24-hour ambulatory blood pressure measurements in children compared to conventional cuff-based devices, while eliminating cuff-induced nocturnal arousals.
PURPOSE: Ambulatory, cuff-less blood pressure (BP) measurement devices are a promising trend to alleviate the strains of conventional, cuff-based BP determination. Cuff-less devices circumvent discomfort and nocturnal arousal reactions which can be triggered by cuff inflation from conventional, cuff-based ambulatory blood pressure measurement devices. Mitigating these discomforts is especially desirable when performing measurement in children. In this study we want to assess the clinical validity of a cuff-less BP measurement device for 24-h measurements in children and adolescents. MATERIALS AND METHODS: We compared the simultaneously retrieved BP data of the cuff-less SOMNOtouch NIBP and the cuff-based Mobil-O-Graph in 24-h use in 90 children in the range from 5 to 17 years old. RESULTS: A total of 1218 valid measurement pairs showed a mean deviation of 0.99 mmHg (limits of agreement: 21.44/-19.46) for systolic and 3.03 mmHg (limits of agreement: 24.37/-18.31) for diastolic BP values. Patient-specific difference of means was within 15 mmHg in 97.7% (systolic BP) and 93.2% (diastolic BP) patients. 25.6% of nocturnal cuff inflations led to determinable, BP-relevant arousal reactions. CONCLUSIONS: The SOMNOtouch NIBP demonstrated little measurement deviation of mean BP compared to the cuff-based technique over a broad spectrum of 24-h, ambulatory BP measurements in children and adolescents. Cuff-less blood pressure measurement relieves the issue of nocturnal arousal reactions which are shown to be frequently induced by cuff-based measurements. Driven by these promising results, we encourage ongoing efforts to create enough evidence on cuff-less BP measurement to promote it into broad clinical application.
No takes yet. Share an insight, caveat, or question.
Bothe et al. (2023) conducted an observational in Blood pressure measurement (n=90). Cuff-less SOMNOtouch NIBP vs. Cuff-based Mobil-O-Graph was evaluated on Mean deviation of systolic and diastolic BP values (MD 0.99 mmHg (systolic) and 3.03 mmHg (diastolic)). The cuff-less SOMNOtouch NIBP demonstrated a mean deviation of 0.99 mmHg for systolic and 3.03 mmHg for diastolic BP compared to a cuff-based device during 24-hour ambulatory measurements.