Key result
A novel invasive method using intravascular ultrasound and a Doppler flow wire to assess peripheral microvascular vasomotor function demonstrated excellent 6-month reproducibility (ICC 0.827-0.995) and concurrent validity.
Why the study?
A novel invasive method using IVUS and a Doppler flow wire was developed to assess peripheral microvascular vasomotor function, but its reproducibility and validity needed evaluation.
Observational (n=49)
Single-blind
No
Novel IVUS-Doppler method assesses profunda microvascular function; leaves open clinical adoption pending validation trials.
In healthy arteries, blood flow is regulated by microvascular tone assessed by changes in blood flow volume and vascular resistance to endothelium-dependent and -independent vasodilators. We developed a novel method of using intravascular ultrasound (IVUS) and a Doppler flow wire to measure changes in blood flow volume and vascular resistance of the profunda arterial bed. We assessed the variability over 6 months in measuring microvascular endothelium-dependent dilation to acetylcholine and endothelium-independent dilation to adenosine in 20 subjects who were part of a larger study of Gulf War Illness without obstructive peripheral artery disease. Vasomotor function was assessed by Infusions of control (dextrose), acetylcholine (10-6M), adenosine (50μg), and nitroglycerin (25μg/ml). 400 IVUS and 240 flow velocity images were measured a mean 6 (SD = 2) months apart blind to measurement and infusion stage. The mean (SD) baseline profunda flow was 227 (172) ml/min and vascular resistance 4.6 x 104 (2.4 x 104) dynes-s/cm5. The intraclass correlation coefficients for 6-month variability for vascular function were excellent (range 0.827-0.995). Bland-Altman analyses showed mean differences of less than 2% for microvascular endothelium-dependent function (flow volume and resistance) and less than 1% for macrovascular endothelium-dependent function with acceptable limits of agreement. In 49 subjects assessing concurrent validity of the technique against atherosclerosis risk factors, we observed greater impairment in microvascular endothelium-dependent function per year of age (flow volume = -1.4% (p = 0.018), vascular resistance = 1.5% (p = 0.015)) and current smoking (flow volume = -36.7% (p = .006), vascular resistance = 50.0% (p<0.001)). This novel method of assessing microvascular vasomotor function had acceptable measurement reproducibility and validity.
No takes yet. Share an insight, caveat, or question.
Kinlay et al. (2019) conducted an observational in Gulf War Illness without obstructive peripheral artery disease (n=49). Intravascular ultrasound (IVUS) and Doppler flow wire assessment was evaluated on 6-month reproducibility of vascular function measurements (intraclass correlation coefficients) (ICC 0.827-0.995). A novel invasive method using intravascular ultrasound and a Doppler flow wire to assess peripheral microvascular vasomotor function demonstrated excellent 6-month reproducibility (ICC 0.827-0.995) and concurrent validity.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:
More from PLoS ONE