Key result
The Vasotens oscillometric BP system showed satisfactory accordance with the SphygmoCor device for central systolic BP and PWV.
Why the study?
Does the Vasotens oscillometric ambulatory BP monitoring system accurately measure central BP, aortic augmentation index, and PWV compared to the standard SphygmoCor device in individuals?
Cross-Sectional (n=99)
Does the Vasotens oscillometric ambulatory BP monitoring system accurately measure central BP, aortic augmentation index, and PWV compared to the standard SphygmoCor device in individuals?
The integration of Vasotens oscillometric technology into an ambulatory BP monitoring system provides accurate measurements of central BP and arterial stiffness indices compared to the standard tonometric method.
Answers are grounded in this paper and cite their sources.
May enable routine central BP and PWV assessment via ABPM; leaves open outcome validation before practice adoption.
BACKGROUND: The objective of this study was to validate the novel integration of oscillometric (Vasotens(®)) technology into a BPLab(®) ambulatory blood pressure (BP) monitoring system to measure central BP, the aortic augmentation index, and pulse wave velocity (PWV) compared with the recommended and widely accepted tonometric method. METHODS: The ARTERY Society guidelines for comparison of PWV measurement techniques were used as the basis for recruitment of 99 individuals (mean age 44±19 years, 52 males). The standard for comparison was the conventional "classic" SphygmoCor device. RESULTS: Accordance of the two methods was satisfactory (r=0.98, mean difference of 2.9±3.5 mmHg for central systolic BP; r=0.98, mean difference of -1.1±2.3 mmHg for central diastolic BP; r=0.83, mean difference of -2.6%±13% for aortic augmentation index; r=0.85, mean difference of 0.69±1.4 for PWV). CONCLUSION: The performance of Vasotens algorithms using an oscillometric ambulatory BP monitoring system is feasible for accurate diagnosis, risk assessment, and evaluation of the effects of antihypertensive drugs.
No takes yet. Share an insight, caveat, or question.
Orlov et al. (2014) reported a cross-sectional. Vasotens technology in BPLab ambulatory BP monitoring system vs. SphygmoCor device was evaluated on central systolic BP (mean difference 2.9±3.5 mmHg). The Vasotens oscillometric BP system showed satisfactory accordance with the SphygmoCor device for central systolic BP and PWV.