Key result
The STENTYS Xposition S balloon delivery system was successfully placed in 100% of cases with no longitudinal geographic miss and a 4% rate of 30-day major adverse cardiac events.
Why the study?
Does the STENTYS Xposition S balloon delivery system allow precise deployment without longitudinal geographic miss in patients with coronary artery disease?
Does the STENTYS Xposition S balloon delivery system allow precise deployment without longitudinal geographic miss in patients with coronary artery disease?
The novel STENTYS Xposition S balloon delivery system is feasible and achieves high technical success without longitudinal geographic miss in patients with coronary artery disease.
AIMS: A novel balloon delivery system (BDS) for the self-apposing STENTYS sirolimus-eluting stent (SES) has been developed for highly precise longitudinal stent positioning and deployment. The aim of this first-in-man study is to report the quantitative coronary analysis (QCA) angiography and optical coherence tomography (OCT) results as well as the 30-day clinical outcomes of the STENTYS Xposition S SES. METHODS AND RESULTS: We included 25 patients (mean age 66.1±10.7 years) with stable coronary artery disease (24%) or acute coronary syndrome (including STEMI in 40%). The device was successfully placed at the intended site in all cases (100%), without procedural complications. Longitudinal geographic miss (entire lesion length [on QCA] not completely covered by the stent) was not observed. Pre-procedural MLD on QCA was 1.30±0.74 mm and post-procedural MLD was 2.74±0.44 mm, p<0.001 (acute gain 1.44±0.70 mm). OCT analyses showed a low percentage of malapposed stent struts directly post stent placement (2.4%), which further decreased after post-dilatation (0.6%, p=0.013), while mean stent area increased (from 9.7 mm2 to 10.5 mm2, p<0.001). At 30-day clinical follow-up, one (4%) major adverse cardiac event (MACE) was observed. One acute stent thrombosis (ST) occurred immediately post procedure in a STEMI patient which was related to inadequate medication therapy. CONCLUSIONS: This first-in-man study demonstrated that the use of the novel STENTYS Xposition S balloon delivery system is feasible with a high technical success rate without longitudinal geographical miss. Stent strut malapposition rate directly after STENTYS placement was low and improved further after post-dilatation.
No takes yet. Share an insight, caveat, or question.
Lu et al. (2016) studied Stable coronary artery disease or acute coronary syndrome (n=25). STENTYS Xposition S balloon delivery system was evaluated on Longitudinal geographic miss. The STENTYS Xposition S balloon delivery system was successfully placed in 100% of cases with no longitudinal geographic miss and a 4% rate of 30-day major adverse cardiac events.