Key result
STENTYS self-apposing stents yield an ~11% two-year MACE rate in STEMI patients.
Why the study?
No long-term data were available on the clinical outcomes of the STENTYS self-apposing stent in STEMI patients undergoing primary PCI.
Observational (n=965)
Yes
The STENTYS self-apposing coronary stent demonstrated feasible long-term outcomes with an 11.2% MACE rate at two years in STEMI patients, with post-dilatation significantly reducing target lesion revascularization and stent thrombosis.
May support self-apposing stent feasibility in STEMI; leaves open randomized comparison to conventional platforms.
AIMS: The APPOSITION III registry evaluated the feasibility and performance of the STENTYS self-apposing stent in an ST-segment elevation myocardial infarction (STEMI) population. This novel self-apposing stent device lowers stent strut malapposition rates and therefore carries the potential to prevent stent undersizing during primary percutaneous coronary intervention (PCI) in STEMI patients. To date, no long-term data are available using this device in the setting of STEMI. We aimed to evaluate the long-term clinical outcomes of the APPOSITION III registry. METHODS AND RESULTS: This was an international, prospective, multicentre post-marketing registry. The study population consisted of 965 STEMI patients. The primary endpoint, major adverse cardiac events (MACE), was defined as the composite of cardiac death, recurrent target vessel myocardial infarction (TV-MI), and clinically driven target lesion revascularisation (CD-TLR). At two years, MACE occurred in 11.2%, cardiac death occurred in 2.3%, TV-MI occurred in 2.3% and CD-TLR in 9.2% of patients. The two-year definite stent thrombosis (ST) rate was 3.3%. Incremental event rates between one- and two-year follow-up were 1.0% for TV-MI, 1.8% for CD-TLR, and 0.5% for definite ST. Post-dilation resulted in significantly reduced CD-TLR and ST rates at 30-day landmark analyses. Results were equivalent between the BMS and PES STENTYS subgroups. CONCLUSIONS: This registry revealed low rates of adverse events at two-year follow-up, with an incremental ST rate as low as 0.5% in the second year, demonstrating that the self-apposing technique is feasible in STEMI patients on long-term follow-up while using post-dilatation.
No takes yet. Share an insight, caveat, or question.
Lu et al. (2017) conducted an observational in ST-segment elevation myocardial infarction (STEMI) (n=965). STENTYS self-apposing coronary stent was evaluated on Major adverse cardiac events (MACE), defined as the composite of cardiac death, recurrent target vessel myocardial infarction (TV-MI), and clinically driven target lesion revascularisation (CD-TLR). The STENTYS self-apposing coronary stent demonstrated a two-year MACE rate of 11.2% and a definite stent thrombosis rate of 3.3% in patients presenting with STEMI.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: