Key result
Central systolic blood pressure estimation based on the QKD interval was comparable in precision to the SphygmoCor device, with an average difference of 2 ± 6 mm Hg.
Why the study?
Does central SBP estimation based on the QKD interval provide comparable precision to the SphygmoCor device in patients?
Observational (n=172)
Does central SBP estimation based on the QKD interval provide comparable precision to the SphygmoCor device in patients?
Answers are grounded in this paper and cite their sources.
Central systolic blood pressure estimation based on the QKD interval provides comparable precision to the SphygmoCor device, offering a potential method for 24-hour ambulatory central blood pressure monitoring.
OBJECTIVE: We validated against invasive measurements a measure of central SBP based on the QKD interval. The present objective is a comparison to the SphygmoCor device. MATERIALS AND METHODS: A first cohort was used to redefine an algorithm for estimating ascending aorta SBP (cSBP) with respect to SphygmoCor. This algorithm was then tested in a second cohort under static conditions and in a third cohort during a head-up tilt to study blood pressure variations. RESULTS: The first cohort (36 patients, mean age: 46 ± 20 years, SBP: 114 ± 19 mm Hg) defined the estimation algorithm as cSBP =62.687+1.35 × meanBP-0.207 × heart rate-0.251 × height-0.1 × QKD (mean difference with SphygmoCor=0 ± 7 mmHg).In the second cohort (105 patients, mean age: 51 ± 19 years, SBP: 132 ± 22 mmHg), the average difference between the two techniques was 2 ± 6 mm Hg (R=0.93). The dynamic cohort included 31 patients (mean age: 49 ± 23 years, SBP: 110 ± 19 mmHg) with 224 pairs of measurement. We studied the changes in central blood pressures compared with the basal state of each of the two techniques during head-up tilt. With QKD, the average cSBP was 104 ± 19 mmHg, the average variation was -13.00 mmHg ± 15 (range 7-74 mmHg). With SphygmoCor, average cSBP was 10 mmHg ± 18, the average variation was -14 mmHg ± 16 (range 2-81 mmHg). The variations in blood pressure with the two techniques were well correlated (R=0.93). CONCLUSION: cSBP estimation based on the QKD interval is comparable in precision to the SphygmoCor. Its ability to measure cSBP variations suggests that it could be employed for ambulatory measurements over 24 h.
No takes yet. Share an insight, caveat, or question.
Crémer et al. (2013) reported an observational. Central SBP estimation based on the QKD interval vs. SphygmoCor device was evaluated on Average difference in central SBP between QKD method and SphygmoCor (MD 2). Central systolic blood pressure estimation based on the QKD interval was comparable in precision to the SphygmoCor device, with an average difference of 2 ± 6 mm Hg.