Key result
Point-of-care capillary lipid testing closely matches standard venous sampling, showing ~5% difference in HDL-C.
Why the study?
Does capillary lipid profiling using the Cobas b101 POCT analyzer provide comparable results to central laboratory venous testing for CVD risk assessment in screened participants?
Cross-Sectional (n=84)
Does capillary lipid profiling using the Cobas b101 POCT analyzer provide comparable results to central laboratory venous testing for CVD risk assessment in screened participants?
Answers are grounded in this paper and cite their sources.
The Cobas b101 POCT analyzer using capillary blood provides reliable non-HDL-C and SCORE2 risk assessments for population-level screening, though triglyceride measurements are less accurate.
Objectives Cardiovascular disease (CVD) is a leading cause of mortality worldwide. Emerging lung cancer screening in smokers may offer an opportunity for simultaneous cardiovascular risk evaluation. We developed a mobile CT screening unit and assessed the feasibility of integrating dyslipidemia screening and non-HDL cholesterol-based SCORE2 estimation using the Cobas ® b101 point-of-care testing (POCT) analyzer on capillary blood samples. Methods Following a preliminary evaluation of two POCT analyzers, the Cobas ® b101 was selected for capillary lipid profiling. Lipid results obtained from capillary samples analyzed in the mobile unit were compared with venous samples analyzed in the central laboratory in 84 screened participants. Results For total cholesterol and HDL-C, the Passing–Bablok regression were strong, with clinically acceptable differences of −0.81 % and +5.23 %, respectively. Triglycerides showed a clinically unacceptable difference of +29.4 %, but strong agreement with clinical categories was observed (kappa 95 % CI [0.55; 0.82]); this may reflect both matrix effects and the non-fasting status of almost all participants (∼91 % were non-fasting). Calculated LDL-C misclassification rates were 47 %, 37 %, and 32 % using the Friedewald, Sampson, and Martin–Hopkins equations, respectively. Non-HDL-C showed a clinically acceptable bias of −2.83 %, supporting its use for prevention goals and familial hypercholesterolemia screening. Concordance analysis of SCORE2 based on non-HDL-C categories revealed high reliability (95 % CI [0.72; 0.92]). Conclusions These findings support the use of Cobas ® b101 with capillary sampling for population-level screening and preliminary CVD risk assessment using SCORE2, despite reduced analytical performance for TG. Abnormal results should be confirmed by a fasting lipid profile.
No takes yet. Share an insight, caveat, or question.
Villemin et al. (2026) conducted a cross-sectional in Cardiovascular risk evaluation (n=84). Cobas b101 point-of-care testing (POCT) analyzer on capillary blood samples vs. Venous samples analyzed in the central laboratory was evaluated on Difference in lipid profile and SCORE2 estimation. Capillary lipid profiling using the Cobas b101 analyzer showed clinically acceptable differences for total cholesterol and HDL-C compared to venous samples.