Key result
A multiobjective hemodynamic control algorithm demonstrated more robust performance in an in vitro mock circulatory simulator compared with controllers relying on individual indices.
A novel multiobjective control algorithm for ventricular assist devices showed improved robustness in an in vitro mock circulatory loop compared to single-index controllers.
Answers are grounded in this paper and cite their sources.
In vitro VAD control algorithm requires in vivo validation before clinical use; leaves open translation of perfusion-collapse trade-off.
Ventricular assist devices now clinically used for treatment of end-stage heart failure require responsive and reliable hemodynamic control to accommodate the continually changing demands of the body. This is an essential ingredient to maintaining a high quality of life. To satisfy this need, a control algorithm involving a trade-off between optimal perfusion and avoidance of ventricular collapse has been developed. An optimal control strategy has been implemented in vitro that combines two competing indices: representing venous return and prevalence of suction. The former is derived from the first derivative of diastolic flow with speed, and the latter derived from the harmonic spectra of the flow signal. The responsiveness of the controller to change in preload and afterload were evaluated in a mock circulatory simulator using a HeartQuest centrifugal blood pump (CF4b, MedQuest Products, Salt Lake City, UT). To avoid the need for flow sensors, a state estimator was used, based on the back-EMF of the actuator. The multiobjective algorithm has demonstrated more robust performance as compared with controllers relying on individual indices.
No takes yet. Share an insight, caveat, or question.
Gwak et al. (2005) studied end-stage heart failure. Multiobjective hemodynamic control algorithm vs. controllers relying on individual indices was evaluated on responsiveness of the controller to change in preload and afterload. A multiobjective hemodynamic control algorithm demonstrated more robust performance in an in vitro mock circulatory simulator compared with controllers relying on individual indices.
Synapse has enriched 10 related papers. Consider them for comparative context: