Answers are grounded in this paper and cite their sources.
High sensitivity to auroral Hα, resolution in elevation angle along a narrow band from horizon to horizon, and simplicity of components are principal features of a new automatic auroral spectrograph. Sky light reaches an all-sky convex mirror by way of an external curved slot in the vertical plane, whose calculated shape gives a flat image of uniform width. This slit image of adjustable width is photographed from above by a standard 16-mm camera with f/0.95 lens, the light having passed through a blazed grating at minimum deviation. In this arrangement distortions in the noncollimated beam and losses by reflection are kept low, and slit adjustment to 25-A resolution enhances detection of the broadened Hα emission line. Satisfactory sensitivity to major airglow lines, hydrogen, and auroral features is obtained in successive 15-min exposures, representing an approximately fivefold improvement in time resolution over that of the IGY patrol spectrograph, with a consequent loss in spectral resolution.
No takes yet. Share an insight, caveat, or question.
Clark et al. (1959) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: