Key result
A novel mass spectrometric assay successfully measured ACE2 activity using the natural substrate Angiotensin II (km=5 mumol/L) and demonstrated inhibition by MLN 4760 (IC50=3 nmol/L).
A novel, specific, and sensitive mass spectrometric assay was developed for measuring ACE2 activity using its natural endogenous peptide substrate Ang II, enabling rapid screening for ACE2.
Answers are grounded in this paper and cite their sources.
Offers validated ACE2 assay for preclinical models; leaves open clinical diagnostic translation.
A novel assay was developed for evaluation of mouse angiotensin-converting enzyme (ACE) 2 and recombinant human ACE2 (rACE2) activity. Using surface-enhanced laser desorption/ionization time of flight mass spectrometry (MS) with ProteinChip Array technology, ACE1 and ACE2 activity could be measured using natural peptide substrates. Plasma from C57BL/6 mice, kidney from wild-type and ACE2 knockout mice, and rACE2 were used for assay validation. Plasma or tissue extracts were incubated with angiotensin I (Ang I; 1296 m/z) or angiotensin II (Ang II; 1045 m/z). Reaction mixtures were spotted onto the ProteinChips WCX2 and peptides detected using surface-enhanced laser desorption/ionization time of flight MS. MS peaks for the substrates, Ang I and Ang II, and the generated peptides, Ang (1-7) and Ang (1-9), were monitored. The ACE2 inhibitor MLN 4760 (0.01 to 100 micromol/L) significantly inhibited rACE2 activity (IC50=3 nmol/L). Ang II was preferably cleaved by rACE2 (km=5 mumol/L), whereas Ang I was not a good substrate for rACE2. There was no detectable ACE2 activity in plasma. Assay specificity was validated in a model of ACE2 gene deletion. In kidney extract from ACE2-deficient mice, there was no generation of Ang (1-7) from Ang II. However, Ang (1-7) was produced when Ang I was used as a substrate. In conclusion, we developed a specific and sensitive assay for ACE2 activity, which used the natural endogenous peptide substrate Ang II. This approach allows for the rapid screening for ACE2, which has applications in drug testing, high-throughput enzymatic assays, and identification of novel substrates/inhibitors of the renin-angiotensin system.
No takes yet. Share an insight, caveat, or question.
Elased et al. (2006) studied ACE2 activity evaluation. Surface-enhanced laser desorption/ionization time of flight mass spectrometry assay was evaluated on ACE2 activity and substrate cleavage. A novel mass spectrometric assay successfully measured ACE2 activity using the natural substrate Angiotensin II (km=5 mumol/L) and demonstrated inhibition by MLN 4760 (IC50=3 nmol/L).
Synapse has enriched 10 related papers. Consider them for comparative context: