Key result
A simple, wide bandwidth (~1 MHz) biopotential amplifier accurately recorded the time course of pacemaker pulses and allowed highly accurate pulse detection and timing.
A novel, simple, wide-bandwidth biopotential amplifier allows for accurate recording and analysis of pacemaker pulse waveforms, which may improve detection of lead failure or fractured wires.
May aid pacemaker lead troubleshooting; leaves open clinical validation in larger cohorts.
Reliable detection of pacemaker pulses is getting more and more important in electrocardiography (ECG) diagnosis. Many studies recommend ECG amplifiers with higher bandwidth to prevent errors. In the past, few pilot studies showed that analysis of pacemaker pulses waveform can enhance diagnosis (eg, lead failure and fractured wire), but they were carried out with inadequate instrumentations for clinical practice. Typically, pacemaker pulses last hundreds of microseconds, edges of pulses elapse in few microseconds, and amplitude may exhibit large variations from few millivolts to volts. Pulse waveforms change often and depend on pacemaker type and programming. A simple, biopotential amplifier made of a few off-the-shelf components is proposed. The circuit fulfills specifications for biopotential amplifiers and offers a large bandwidth (~1 MHz). Therefore, it is able to accurately record time course of pacemaker pulses and allows highly accurate pulse detection and timing. Signals can be easily displayed and acquired by means of a standard, battery-powered oscilloscope. Pacemaker pulse vectorcardiography can be obtained by using two or more, wideband channels. Some exemplificative waveforms recorded during patient's periodic medical examination are reported. The proposed circuit offers simultaneous conventional ECG signal as an additional output.
No takes yet. Share an insight, caveat, or question.
Cesarelli et al. (2016) studied Pacemaker pulse detection. Wide bandwidth biopotential amplifier was evaluated on Accurate recording of time course of pacemaker pulses and pulse detection. A simple, wide bandwidth (~1 MHz) biopotential amplifier accurately recorded the time course of pacemaker pulses and allowed highly accurate pulse detection and timing.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: