Key result
Percutaneous lead revision for myocardial perforation after CIED placement was successful in 100% of cases (31/31), with all patients surviving without open chest surgery.
Why the study?
Is percutaneous lead revision safe and effective for managing myocardial perforation after CIED placement?
Observational (n=31)
Is percutaneous lead revision safe and effective for managing myocardial perforation after CIED placement?
Percutaneous lead revision for CIED-related myocardial perforation is feasible and effective, and prophylactic pericardial drainage may reduce the risk of urgent pericardiocentesis.
Answers are grounded in this paper and cite their sources.
Supports percutaneous revision for CIED perforation; leaves open need for prospective trials before practice change.
INTRODUCTION: Cardiac perforation is an infrequent but potentially life-threatening complication associated with placement of a cardiac implantable electronic device (CIED). The objective of this study was to determine the outcomes of percutaneous lead revision in patients who had lead perforation of the myocardium after CIED placement. METHODS AND RESULTS: We reviewed records of 1,458 patients who underwent CIED lead extraction or repositioning. Of these, 31 (2.1%) had the procedure performed for lead perforation as a complication of CIED placement. Demographic, clinical, and follow-up characteristics of the patients were analyzed. Mean (SD) patient age was 65 (23) years. Cardiac perforation was detected within 24 hours after implantation in 9 patients, within 1 month in 17, and greater than 1 month in 5. Pericardiocentesis was performed with a pigtail drainage catheter in place before the lead revision in 17 patients (55%) who had pericardial effusion, with or without hemodynamic compromise. All culprit leads were successfully managed with percutaneous lead removal (n = 3 [10%]), new lead placement (n = 12 [38%]), or lead repositioning (n = 16 [52%]). Of the 17 patients with pericardiocentesis before the reoperation, none had tamponade develop; in contrast, 3 of the remaining 14 patients had tamponade develop and required urgent pericardiocentesis. All patients survived without requiring open chest surgery. CONCLUSION: Percutaneous removal or repositioning of the perforating lead is feasible and appears effective. Placement of a prophylactic pericardial drain catheter may reduce the incidence of urgent pericardiocentesis during or after a procedure.
No takes yet. Share an insight, caveat, or question.
Huang et al. (2014) conducted an observational in Myocardial perforation after CIED placement (n=31). Percutaneous lead revision was evaluated on Successful management and survival without requiring open chest surgery. Percutaneous lead revision for myocardial perforation after CIED placement was successful in 100% of cases (31/31), with all patients surviving without open chest surgery.
Synapse has enriched 10 related papers. Consider them for comparative context: