Key result
Single pass lead VDD pacing maintained reliable atrial sensing with the lowest P wave amplitude above 0.2 mV and achieved over 99% atrial synchronous ventricular pacing across different body positions.
Why the study?
Does a single-pass lead VDD pacing system maintain reliable atrial sensing and ventricular pacing across different body positions and physical activities in patients with high degree AV block?
Observational (n=7)
No
Does a single-pass lead VDD pacing system maintain reliable atrial sensing and ventricular pacing across different body positions and physical activities in patients with high degree AV block?
Answers are grounded in this paper and cite their sources.
Single-pass lead VDD pacing provides reliable atrial sensing and atrioventricular synchrony regardless of body position or physical activity in patients with high-degree AV block and intact sinus node function.
BackgroundSingle pass lead VDD pacing preserves atrioventricular synchrony with a single lead system which incorporates floating atrial electrodes.The objectives of this study were to measure whether different body postures and physical activities cause significant changes of the atrial electrogram amplitudes and to evaluate the effectiveness of its atrial sensing, ventricular sensing and pacing.MethodProspective study was done in 7 patients with high degree AV block and normal sinus node function in whom a single lead VDD pacing system was implanted.The P wave amplitude was been measured in different condition during follow-up period.Results1) During follow-up period, the P wave amplitude showed variation with changes in posture and respiration, but there was no consistent increase or decrease in amplitude.The lowest P wave amplitude was above the minimal atrial sensing value of 0.2 mV. 2) The percentage of atrial synchronous ventricular pacing recorded in Holter ECG and during Treadmill exercise test was more than 99%.3) Atrial oversensing or VA cross sensing were not observed in any of the patients.ConclusionsDespite floating atrial electrode, the single pass lead VDD pacing maintains reliable atrial sensing and ventricular pacing in different body position and physical activity, so it may offer an excellent alternative in patients with high grade AV block and intact sinus node function.( ( ( (Korean Circulation J 1999 ; 29( ( ( (10) ) ) ) : 1082-1088) ) ) ) KEY WORDSSingle pass leadVDD pacingP wave amplitudeAtrial synchronous ventricular pacing. Dual chamber 1-4) VVI .
No takes yet. Share an insight, caveat, or question.
Kang et al. (1999) conducted an observational in High degree AV block and normal sinus node function (n=7). Single pass lead VDD pacing was evaluated on Percentage of atrial synchronous ventricular pacing. Single pass lead VDD pacing maintained reliable atrial sensing with the lowest P wave amplitude above 0.2 mV and achieved over 99% atrial synchronous ventricular pacing across different body positions.