Key result
A 62-year-old man with a CRT-D developed reel syndrome due to spontaneous rotation of the unanchored generator, resulting in left ventricular lead displacement and worsening dyspnoea.
Case Report (n=1)
Reel syndrome, caused by the rotation of an unsecured pacemaker generator on its transverse axis, is a potential cause of lead displacement and CRT non-response.
Generator migration may occur without pectoral fixation; case report leaves open whether routine suturing prevents dislodgement in CRT-D patients.
A 62-year-old man with ischaemic cardiomyopathy complained of an increase of basal dyspnoea. A CRT-D (Medtronic InSync III Marquis, Minneapolis, MN, USA) was implanted 7 months ago. The leads were inserted without complication through the left subclavian vein and secured with a non-absorbable ligature to the fascia. The generator was not sutured to the pectoral fascia. The patient had been better since the device was implanted, but in the last 3 months had worsened. A radiograph of the chest demonstrated a displacement of the left ventricular lead (Medtronic Attain 4194, 88 cm, Minneapolis, MN, USA), with winding of the lead around the pacemaker box, which presented an horary rotation (Panel A: pos CRT-D implantation; Panel B: left lead dislocation). When he was interrogated he referred a spontaneous twist of the box 3 months ago. Spontaneous displacement of the pacemaker lead due to the rotation of the generator on its transverse axis is known as the ‘reel syndrome’. An insufficient fixation of the leads/generator may contribute to it. In this case, the rotation of the generator (not secured to the fascia) resulted in reeling in of the pacemaker lead. The box and a new lead were fixed securely and carefully without complications. Conflict of interest: none declared.
No takes yet. Share an insight, caveat, or question.
López et al. (2010) conducted a case report in Ischaemic cardiomyopathy (n=1). CRT-D implantation was evaluated. A 62-year-old man with a CRT-D developed reel syndrome due to spontaneous rotation of the unanchored generator, resulting in left ventricular lead displacement and worsening dyspnoea.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: