Key result
Real-time three-dimensional transesophageal echocardiography demonstrated better agreement with the implanted left atrial appendage occluder size compared to two-dimensional echocardiography (weighted kappa 0.62 vs. 0.28).
Why the study?
Does RT3D TEE improve measurement reproducibility and device size selection agreement compared to RT2D TEE in patients undergoing LAAO?
Cross-Sectional (n=40)
No
Does RT3D TEE improve measurement reproducibility and device size selection agreement compared to RT2D TEE in patients undergoing LAAO?
Absolute Event Rate: 0.62% vs 0.28%
RT3D TEE provides more reproducible measurements of the left atrial appendage and better agreement with final implanted device size compared to RT2D TEE during LAA occlusion.
May aid LAAO sizing decisions; leaves open need for prospective outcome trials.
BACKGROUND: Real-time three-dimensional transesophageal echocardiography (RT3D TEE) enables better visualization of the left atrial appendage (LAA) and may be superior to real-time two-dimensional transesophageal echocardiography (RT2D TEE) for LAA occlusion (LAAO). The aim of this study was to assess inter- and intra-observer variability of RT2D TEE and RT3D TEE measurements of LAA, and to assess the accordance of RT2D TEE and RT3D TEE with appropriate occluder selection. METHODS: Transesophageal echocardiography was performed in 40 patients during LAAO. RT2D TEE and RT3D TEE measurements of the ostium and landing zone were performed independently by two echocardiographers. The appropriate choice of occluder was confirmed with fluoroscopic criteria. After the procedures, RT2D TEE and RT3D TEE evaluation were repeated separately by the same echocardiographers. RESULTS: The mean ostium diameters by RT2D TEE obtained by the two observers were 23.6 ± 4.2 vs. 24.8 ± 5.2 (p = 0.04), and the mean landing zone diameters were 17.7 ± 4.4 vs. 19.4 ± 3.9 (p < 0.01). In the case of RT3D TEE, the ostium diameters were 29.6 ± 5.3 vs. 29.4 ± 6.4 (p = not significant [NS]) and the landing zone diameters were 21.4 ± 3.8 vs. 21.6 ± 3.9 (p = NS). Intra-observer differences were absent in the case of RT3D TEE. The comparison of RT2D TEE vs. RT3D TEE analyses performed by the same echocardiographer revealed significant differences in the ostium and landing zone measurements (both p < 0.01). Agreement between the suggested device size was better for RT3D TEE (weighted kappa was 0.62 vs. 0.28, respectively). CONCLUSIONS: The results obtained with RT3D TEE showed significantly larger dimensions of the ostium and the landing zone. RT3D TEE showed lesser inter- and intra-observer variability and better agreement with the implanted device.
No takes yet. Share an insight, caveat, or question.
Streb et al. (2018) conducted a cross-sectional in Atrial fibrillation (n=40). Real-time three-dimensional transesophageal echocardiography (RT3D TEE) vs. Real-time two-dimensional transesophageal echocardiography (RT2D TEE) was evaluated on Agreement between suggested device size and implanted device size (weighted kappa). Real-time three-dimensional transesophageal echocardiography demonstrated better agreement with the implanted left atrial appendage occluder size compared to two-dimensional echocardiography (weighted kappa 0.62 vs. 0.28).