Key result
Chronic strength duration curves were shifted upward and to the right of acute curves, suggesting that variable pulse width exceeding 1 ms may not significantly reduce stimulation thresholds.
Why the study?
How do chronic strength duration curves compare to acute strength duration curves in patients with cardiac pacemakers?
Observational (n=66)
How do chronic strength duration curves compare to acute strength duration curves in patients with cardiac pacemakers?
Chronic pacemaker stimulation thresholds are shifted upward and to the right compared to acute thresholds, suggesting that increasing pulse width beyond 1 ms may not significantly reduce stimulation thresholds.
Answers are grounded in this paper and cite their sources.
Caution advised against assuming benefit from longer pulse widths in chronic implants; extends acute observations but remains hypothesis-generating.
Successful long-term cardiac pacemaker therapy requires both reliable generator output and stable, attainable chronic stimulation thresholds. Strength duration curves, which display cardiac stimulation thresholds as a function of impulse duration, provide needed data to evaluate more thoroughly acute and chronic lead thresholds. We measured chronic strength duration curves at the time of generator replacement in 47 patients with three different types of lead systems of variable geometry and surface area and compared these curves with acute strength duration curves measured at the time of initial implantation in 19 patients. We found that strength duration curves has a characteristic shape; however, chronic strength duration curves were shifted upward and to the right of acute strength duration curves, regardless of electrode tip surface area or geometric design. Our data suggest that variable pulse width exceeding 1 ms may not significantly reduce stimulation thresholds.
No takes yet. Share an insight, caveat, or question.
Hynes et al. (1981) conducted an observational in Cardiac pacemaker therapy (n=66). Chronic pacemaker leads (at generator replacement) vs. Acute pacemaker leads (at initial implantation) was evaluated on Strength duration curves (stimulation thresholds as a function of impulse duration). Chronic strength duration curves were shifted upward and to the right of acute curves, suggesting that variable pulse width exceeding 1 ms may not significantly reduce stimulation thresholds.
Synapse has enriched 10 related papers. Consider them for comparative context: