Key result
A blood pool subtraction technique enabled correct localization of myocardial infarction at 6 hours postinjection in 88% of patients, compared to 76% using standard 24-hour images.
Why the study?
Does a blood pool subtraction technique allow accurate visualization of myocardial infarction 6 hours postinjection of 111In-antimyosin antibodies in patients with proven MI?
Cross-Sectional (n=25)
Blinded
Does a blood pool subtraction technique allow accurate visualization of myocardial infarction 6 hours postinjection of 111In-antimyosin antibodies in patients with proven MI?
Answers are grounded in this paper and cite their sources.
A blood pool subtraction technique enables accurate visualization of myocardial infarction just 6 hours after 111In-antimyosin antibody injection, potentially accelerating diagnostic management.
111In-antimyosin antibodies are capable of visualizing myocardial infarction (MI). Because of slow blood clearance, images are usually recorded 24 or 48 h postinjection. In this pilot study, a blood pool subtraction technique, which makes it possible to visualize MI 6 h postinjection, is validated. Twenty-five patients with proven MI (16 anterior, 9 inferior) were imaged a few minutes, 6 and 24 h after an injection of 111 MBq 111In-labelled antimyosin antibodies. Three planar views are obtained each time. Using software which performs the geometric registration, the grey level normalization and the subtraction of images, the blood pool image (obtained a few minutes postinjection) is subtracted from the 6 h image. The resulting image is the blood pool corrected 6 h image. The 24 h images and the blood pool corrected 6 h images were interpreted blindly and the number of correct, incorrect and impossible MI localizations was counted. The number of correct localizations is 19/25 for the standard 24 h images and 22/25 for the blood pool corrected 6 h images. Then, with this blood pool subtraction method, it is possible to visualize MI 6 h postinjection. This has to be taken into account when discussing the role of antimyosin scintigraphy in the management of patients with MI.
No takes yet. Share an insight, caveat, or question.
Liehn et al. (1992) conducted a cross-sectional in Myocardial infarction (n=25). Blood pool subtraction technique for 111In-antimyosin antibody imaging vs. Standard 24 h images was evaluated on Correct myocardial infarction localizations. A blood pool subtraction technique enabled correct localization of myocardial infarction at 6 hours postinjection in 88% of patients, compared to 76% using standard 24-hour images.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: