Key result
Rotational intracardiac echocardiography showed moderate agreement with TEE for PFO device selection (72%, κ=0.53), but significant disagreement in patients with atrial septal aneurysm (36%, κ=0.02).
Why the study?
Does rotational intracardiac echocardiography agree with transesophageal echocardiography in measuring fossa ovalis and selecting devices for PFO closure in patients with cryptogenic stroke or peripheral embolism?
Observational (n=45)
Does rotational intracardiac echocardiography agree with transesophageal echocardiography in measuring fossa ovalis and selecting devices for PFO closure in patients with cryptogenic stroke or peripheral embolism?
Effect estimate: κ = 0.53
p-value: p=<0.001
TEE and rotational ICE cannot be used interchangeably for measuring the fossa ovalis and selecting devices for PFO closure, particularly in patients with atrial septal aneurysm.
May warrant caution in relying on rotational ICE for PFO device sizing; leaves open which modality best predicts closure success.
BACKGROUND: Transesophageal (TEE) and intracardiac (ICE) echocardiography are commonly used to guide percutaneous patent foramen ovale (PFO) closure. The study aim was to perform a head-to-head comparison between TEE and rotational ICE echocardiography in the measurement of the fossa ovalis and device selection. METHODS: In 45 patients with cryptogenic stroke or peripheral embolism and PFO with large right-to-left shunt, fossa ovalis dimensions were assessed preoperatively by TEE and intraoperatively by rotational ICE. The Amplatzer devices, deployed on the basis of ICE, were compared with those that would have been selected by TEE. RESULTS: A good correlation between TEE and rotational ICE was observed for both longitudinal and transverse fossa ovalis dimensions (TEE four-chamber vs. ICE four-chamber: r = 0.75; TEE bicaval vs. ICE four-chamber: r = 0.77; TEE aorta vs. ICE aorta: r = 0.59; P < 0.001 for all). However, no such correlation was found in 13 patients with atrial septal aneurysm (ASA) (TEE four-chamber vs. ICE four-chamber: r = 0.33; TEE bicaval vs. ICE four-chamber: r = 0.49; TEE aorta vs. ICE aorta: r = 0.05; P = NS for all). At Bland-Altman analysis, slight systematic differences with wide limits of agreement for each comparison were observed, particularly in patients with ASA, suggesting that the two imaging modalities cannot be used interchangeably. As regards device selection, a moderate agreement was found between TEE- and ICE-guided device size (72%, κ = 0.53, P < 0.001), except in patients with ASA (36%, κ = 0.02, P = NS). CONCLUSIONS: Our study suggests a significant disagreement between TEE and rotational ICE in measuring fossa ovalis and selecting the device for PFO closure, particularly in patients with ASA.
No takes yet. Share an insight, caveat, or question.
Vigna et al. (2012) conducted an observational in Cryptogenic stroke or peripheral embolism and PFO with large right-to-left shunt (n=45). Rotational intracardiac echocardiography (ICE) vs. Transesophageal echocardiography (TEE) was evaluated on Measurement of the fossa ovalis and device selection (κ = 0.53, p=<0.001). Rotational intracardiac echocardiography showed moderate agreement with TEE for PFO device selection (72%, κ=0.53), but significant disagreement in patients with atrial septal aneurysm (36%, κ=0.02).