Key result
The perimeter-derived diameter on cardiac CT images provided the most accurate estimation of LAAO device size with a mean error of -0.8 mm, regardless of landing zone eccentricity.
Why the study?
Data were lacking on which cardiac CT-derived parameter to use during pre-procedural planning of LAAO, given limitations of two-dimensional transesophageal echocardiography for evaluating complex LAA anatomy.
Does perimeter-derived diameter from cardiac CT improve the accuracy of LAAO device size prediction compared to other CT parameters in patients undergoing LAAO?
Observational (n=62)
Yes
Does perimeter-derived diameter from cardiac CT improve the accuracy of LAAO device size prediction compared to other CT parameters in patients undergoing LAAO?
p-value: p=<0.05
Perimeter-derived diameter from cardiac CT provides the most accurate estimation of LAAO device size regardless of landing zone eccentricity, rendering oversizing unnecessary.
Supports perimeter-derived CT diameter for LAAO sizing accuracy; leaves open prospective outcome impact.
The complex structure of the left atrial appendage (LAA) brings limitations to the two-dimensional-based LAA occlusion (LAAO) size prediction system using transesophageal echocardiography. The LAA anatomy can be evaluated more precisely using three-dimensional images from cardiac computed tomography (CT); however, there is lack of data regarding which parameter to choose from CT-based images during pre-procedural planning of LAAO. We aimed to assess the accuracy of measurements derived from cardiac CT images for selecting LAAO devices. We retrospectively reviewed 62 patients with Amplatzer Cardiac Plug and Amulet LAAO devices who underwent implantation from 2017 to 2020. The minimal, maximal, average, area-derived, and perimeter-derived diameters of the LAA landing zone were measured using CT-based images. Predicted device sizes using sizing charts were compared with actual successfully implanted device sizes. The mean size of implanted devices was 27.1 ± 3.7 mm. The perimeter-derived diameter predicted device size most accurately (mean error = - 0.8 ± 2.4 mm). All other parameters showed significantly larger error (mean error; minimal diameter = - 4.9 ± 3.3 mm, maximal diameter = 1.0 ± 2.9 mm, average diameter = - 1.6 ± 2.6 mm, area-derived diameter = - 2.0 ± 2.6 mm) than the perimeter-derived diameter (all p for difference < 0.05). The error for other parameters were larger in cases with more eccentrically-shaped landing zones, while the perimeter-derived diameter had minor error regardless of eccentricity. When oversizing was used, all parameters showed significant disagreement. The perimeter-derived diameter on cardiac CT images provided the most accurate estimation of LAAO device size regardless of landing zone eccentricity. Oversizing was unnecessary when using cardiac CT to predict an accurate LAAO size.
No takes yet. Share an insight, caveat, or question.
Cho et al. (2021) conducted an observational in Atrial fibrillation (n=62). Perimeter-derived diameter on 3D cardiac CT vs. Other CT-derived diameters (minimal, maximal, average, area-derived) was evaluated on Mean error of predicted device size compared to actual implanted device size (p=<0.05). The perimeter-derived diameter on cardiac CT images provided the most accurate estimation of LAAO device size with a mean error of -0.8 mm, regardless of landing zone eccentricity.