Key result
Pressure half time was closely correlated with the diastolic time interval across heart rates of 75 to 150 beats/min in patients with normally functioning mitral valve prostheses.
Why the study?
Does changing heart rate affect pressure half time in patients with normally functioning mitral valve prostheses?
Observational (n=8)
Does changing heart rate affect pressure half time in patients with normally functioning mitral valve prostheses?
In normally functioning prosthetic mitral valves, pressure half time is significantly influenced by the diastolic time interval and heart rate, which may obscure the assessment of orifice area.
Answers are grounded in this paper and cite their sources.
Pressure half-time requires heart-rate adjustment in mitral prostheses; leaves validation of clinical thresholds open.
To test the validity of a relation between the pressure half time and the diastolic time interval, previously shown in a pulse duplication system, eight patients with prosthetic mitral valves and permanent pacemaker systems were studied. Recordings were made from the apex by continuous wave or pulsed Doppler echocardiography at heart rates between 75 and 150 beats/min. The pressure half time was found to be closely correlated with the diastolic time interval although there was individual variation and in three prostheses the pressure half time attained a plateau when the diastolic time interval was more than 300 ms. It is likely that the orifice area is the main controller of pressure half time where there is stenosis of the prosthesis, but that other factors such as ventricular or atrial compliance and the diastolic time interval may modify or obscure the effect of orifice area in normally functioning prosthetic valves.
No takes yet. Share an insight, caveat, or question.
Chambers et al. (1988) conducted an observational in Prosthetic mitral valves and permanent pacemaker systems (n=8). Changes in heart rate (75 to 150 beats/min) was evaluated on Correlation between pressure half time and diastolic time interval. Pressure half time was closely correlated with the diastolic time interval across heart rates of 75 to 150 beats/min in patients with normally functioning mitral valve prostheses.