Key result
Echocardiography and phonocardiography of the Lillehei-Kaster mitral valve prosthesis demonstrated a disc excursion of 7-12 mm, mean opening velocity of 37.7 cm/sec, and a two-fold closing sound.
Observational (n=22)
Establishing the normal echocardiographic and phonocardiographic profile of the Lillehei-Kaster mitral valve prosthesis aids in the noninvasive evaluation of its function and the detection of complications.
Answers are grounded in this paper and cite their sources.
Normal L-K mitral prosthesis shows stenosis-like ECHO with defined velocities; leaves open differentiation from dysfunction.
The Lillehei-Kaster (L-K) valve is a tilting disc prosthesis currently in use for heart valve replacement. We report data from phonocardiography (PHONO) and echocardiography (ECHO) in 20 patients with a normally functioning mitral valve prosthesis (MVP). The MVP is well recorded by ECHO, resembling a mitral stenosis pattern with a disc excursion of 7 to 12 mm and a mean opening and closing velocity of 37.7 and 59.8 cm/sec, respectively. Combined PHONO and ECHO show that the opening sound of the MVP is small, related to onset of valve opening and not peak opening. It follows the aortic component of the second sound (A 2 ) by 0.05 to 0.09 sec in normals; it may be absent (12/20). Peak opening follows A 2 by 0.08 to 0.12 sec. The closing sound of the MVP, best heard in the mitral area, is always two-fold with a small initial high frequency (HF) component and a second large HF sound separated by no more than 0.03 sec. The first component (A) is related to onset of closure and the second (B) to completion of closure. Systolic and diastolic murmurs were commonly present. These data establish the normal ECHO and PHONO findings for patients with L-K MVP. Two other patients are also reported in whom such data were helpful for the noninvasive evaluation of valve function, indicating in one instance acute mitral regurgitation and in the other an increase in degree of aortic regurgitation.
No takes yet. Share an insight, caveat, or question.
Gibson et al. (1974) conducted an observational in Lillehei-Kaster mitral valve prosthesis (n=22). Echocardiography and phonocardiography of Lillehei-Kaster mitral valve prosthesis was evaluated on Echocardiographic and phonocardiographic characteristics (disc excursion, opening/closing velocity, timing of sounds). Echocardiography and phonocardiography of the Lillehei-Kaster mitral valve prosthesis demonstrated a disc excursion of 7-12 mm, mean opening velocity of 37.7 cm/sec, and a two-fold closing sound.
Synapse has enriched 10 related papers. Consider them for comparative context: