Key result
Abnormal echo tracings in mitral valve prostheses are linked to a ~90% complication rate.
Why the study?
Can echocardiography detect significant abnormalities and predict complications in patients with mitral valve prostheses?
Observational (n=33)
Can echocardiography detect significant abnormalities and predict complications in patients with mitral valve prostheses?
Echocardiography is a valuable non-invasive tool for assessing the function of mitral valve prostheses and predicting prosthesis-related complications.
Answers are grounded in this paper and cite their sources.
May support closer monitoring of mitral prosthesis patients with abnormal tracings; hypothesis-generating and requires prospective validation.
Thirty-three patients with mitral valve prostheses were studied with echocardiography in an effort to determine if this technique could be useful in detecting significant abnormalities. Recordings were obtained in the supine position with the transducer directed to record maximum excursion of the prosthesis. Echoes from the struts, poppet and sewing ring were readily recorded. Amplitude of excursion and opening and closing velocities of the poppet were measured. Fifty echocardiographic recordings were obtained from the 33 patients. Of the 33 patients studied, 22 were thought to have "normal" echo tracings while in 11, the tracings were considered "abnormal." Apparent abnormalities consisted of: 1) abnormal diastolic separation between the poppet and strut, 2) increased echoes near the poppet, strut or sewing ring and 3) a combination of both. There was only one instance of suspected "sticking" of the prosthesis. All patients who had "abnormal" studies except one developed complications associated with their prosthesis (90%) compared to only 36% in patients with "normal" tracings. Five patients in each group died. Autopsy studies are described and correlations with the echocardiographic findings are made. In low profile valves reduction in excursion of the disc may be an indication of malfunction. Echocardiography appears to be of value in the assessment of function of mitral valve prostheses.
No takes yet. Share an insight, caveat, or question.
Watts et al. (1975) conducted an observational in Mitral valve prostheses (n=33). Abnormal echocardiographic tracings vs. Normal echocardiographic tracings was evaluated on Complications associated with prosthesis. Abnormal echocardiographic tracings in patients with mitral valve prostheses were associated with a 90% complication rate compared to 36% in those with normal tracings.
Synapse has enriched 10 related papers. Consider them for comparative context: