Key result
An adaptive noise canceller decreased contaminated ambient noise by about 16 dB from the sound signal of an artificial heart implanted in a goat.
Why the study?
Does an adaptive noise canceller improve denoising of ambient noise from artificial heart sound signals in a goat model?
Does an adaptive noise canceller improve denoising of ambient noise from artificial heart sound signals in a goat model?
An adaptive noise canceller successfully reduced ambient noise by 16 dB in an electro-stethoscope system monitoring an implanted artificial heart in a goat model.
Answers are grounded in this paper and cite their sources.
May improve acoustic monitoring of artificial hearts; leaves open clinical translation from this animal model.
The early diagnosis for malfunction of a mechanical artificial heart implanted in a patient is important to ensure the safety of them. We have developed an electro-stethoscope system that enables malfunction of an artificial heart. However it had a problem that ambient sound interference on artificial heart's sound prevented automatic diagnosis for malfunction of the artificial heart. In this study, we have devised an adaptive noise canceller for efficient denoising of ambient noise from measured sound signal. Artificial heart's sound signal d (n)(=pure artificial heart's sound s (n)+ noise x′ (n)) measured on chest wall above an implanted artificial heart, and ambient noise x (n) measured at body surface are the primary and reference inputs respectively. y (n)(=x (n) h (n)) is the output of a transversal filter in the adaptive noise canceller where h (n) is a filter coefficient vector. The adaptive noise canceller is operated by minimizing the mean-square error E[e2 (n)] with tuning h (n) where e(n) is the output of the noise canceller defined as e (n) = d (n)-y (n). The adaptive noise canceller was evaluated in vivo experiment using a goat implanted an undulation pump ventricular assist device. It has decreased contaminated ambient noise about 16 dB (noise decrease ratio; 42%) from artificial heart's sound signal.
No takes yet. Share an insight, caveat, or question.
Makino et al. (2004) studied Implanted mechanical artificial heart (n=1). Adaptive noise canceller for electro-stethoscope system was evaluated on Decrease in contaminated ambient noise from artificial heart's sound signal. An adaptive noise canceller decreased contaminated ambient noise by about 16 dB from the sound signal of an artificial heart implanted in a goat.
Synapse has enriched 10 related papers. Consider them for comparative context: