Key result
The novel D-shaped bileaflet bioprosthesis demonstrated lower turbulence intensity and greater mean kinetic energy conservation compared to the St. Jude Epic valve, while maintaining similar Reynolds shear stress levels.
Why the study?
Does a novel D-shaped bileaflet porcine bioprosthesis improve physiological left ventricular flow patterns compared to the St. Jude Epic valve in an in-vitro model?
Does a novel D-shaped bileaflet porcine bioprosthesis improve physiological left ventricular flow patterns compared to the St. Jude Epic valve in an in-vitro model?
Answers are grounded in this paper and cite their sources.
A novel biomimetic D-shaped bileaflet bioprosthesis for the mitral position shows promising in-vitro hemodynamic performance by better preserving physiological left ventricular flow patterns.
Prior studies have shown that in a healthy heart, there exist a large asymmetric vortex structure that aids in establishing a steady flow field in the left ventricle. However, the implantation of existing artificial heart valves at the mitral position is found to have a negative effect on this physiological flow pattern. In light of this, a novel D-shaped bileaflet porcine bioprosthesis (GD valve) has been designed based on the native geometry mitral valve, with the hypothesis that biomimicry in valve design can restore physiological left ventricle flow patterns after valve implantation. An in-vitro experiment using two dimensional particle velocimetry imaging was carried out to determine the hemodynamic performance of the new bileaflet design and then compared to that of the well-established St. Jude Epic valve which functioned as a control in the experiment. Although both valves were found to have similar Reynolds shear stress and Turbulent Kinetic Energy levels, the novel D-shape valve was found to have lower turbulence intensity and greater mean kinetic energy conservation.
No takes yet. Share an insight, caveat, or question.
Tan et al. (2016) studied Mitral valve replacement (in-vitro simulation). D-shaped bileaflet porcine bioprosthesis (GD valve) vs. St. Jude Medical Epic Valve was evaluated on Hemodynamic performance (Reynolds shear stress, Turbulent Kinetic Energy, Mean Kinetic Energy, turbulence intensity). The novel D-shaped bileaflet bioprosthesis demonstrated lower turbulence intensity and greater mean kinetic energy conservation compared to the St. Jude Epic valve, while maintaining similar Reynolds shear stress levels.