Key result
Capillarys 2 Flex Piercing demonstrated strong analytical performance for HbA1c quantification, with a mean difference of -0.03 (95% CI -0.05 to -0.0003) and strong correlation (r > 0.99).
Population
Blood samples for HbA1c measurement
Comparison
Capillarys 2 Flex Piercing using Capillary… vs HPLC
Design
Other
Authors
Loading...
May support HbA1c assay adoption in labs; leaves open outcome correlation studies.
Mean Difference: -0.03 (95% CI -0.05–-0.0003)
The Capillarys 2 Flex Piercing system provides reliable analytical performance for HbA1c quantification, even at low total hemoglobin levels, and allows simultaneous HbA2 quantification.
Doggui et al. (2017) studied Diabetes. Capillarys 2 Flex Piercing (C2FP) vs. HPLC was evaluated on HbA1c measurement method comparison (MD -0.03, 95% CI -0.05 to -0.0003). Capillarys 2 Flex Piercing demonstrated strong analytical performance for HbA1c quantification, with a mean difference of -0.03 (95% CI -0.05 to -0.0003) and strong correlation (r > 0.99).