Key result
Telemetry-derived 7-lead and single-lead ECGs underestimated the corrected QT interval by 13.45±32.05 msec and 19.62±33.19 msec compared to standard 12-lead ECGs (p<0.05).
Why the study?
Severely ill inpatients with COVID-19 require close ECG monitoring due to cardiac involvement and drug side effects, raising the question of whether telemetry-generated ECGs could serve as an alternative to conventional 12-lead ECGs.
Does telemetry-generated 7-lead or single-lead ECG accurately measure ECG parameters compared to conventional 12-lead ECG in ICU patients with COVID-19?
Observational (n=33)
Does telemetry-generated 7-lead or single-lead ECG accurately measure ECG parameters compared to conventional 12-lead ECG in ICU patients with COVID-19?
Mean Difference: -13.45
p-value: p=< 0.05
Telemetry-derived ECGs underestimate QT intervals and overestimate ST/T wave changes compared to standard 12-lead ECGs, indicating they are not a reliable replacement for routine diagnosis in critically ill patients.
Introduction: Severely ill inpatients with SARS-CoV-2 infection, Coronavirus Disease 2019 (COVID-19) require close electrocardiographic (ECG) monitoring due to frequent cardiac involvement of the disease and cardiovascular side effects of therapies. This study aimed to compare ECG parameters measured from conventional 12-lead ECGs to those from a telemetry-generated 7-lead or single lead ECG to determine if the latter may be an alternative for screening and monitoring patients, particularly during a pandemic. Methods and Results: We identified 33 patients with respiratory failure due to COVID-19 undergoing telemetry monitoring in the intensive care unit. Each received a 12-lead ECG utilizing standard lead placement. A concurrent 7-lead ECG and single lead (lead II) tracing were obtained using the central telemetry system. Each ECG was interpreted and intervals manually measured by 2 cardiologists with disagreements adjudicated by a third. Compared to the 12-lead ECG measurement, the 7-lead ECG underestimated the corrected QT by on average 13.45±32.05 msec, and the single lead ECG underestimated corrected QT by 19.62±33.19 msec (Bazett, p < 0.05). Bland Altman analysis also demonstrated evidence of a positive bias, suggesting that the telemetry-derived tracings underestimated the QT interval. The presence of T wave abnormalities and ST segment changes were overestimated by the telemetry-derived tracings as compared to standard ECGs. Conclusion: Though telemetry-derived ECGs may be useful in screening patients for significant ECG abnormalities, they likely do not represent a reliable replacement of the standard 12-lead ECG in the routine diagnosis and management of critically ill patients.
No takes yet. Share an insight, caveat, or question.
Coromilas et al. (2020) conducted an observational in Respiratory failure due to COVID-19 (n=33). Telemetry-generated 7-lead or single lead ECG vs. Conventional 12-lead ECG was evaluated on Corrected QT interval (MD -13.45 msec, p=< 0.05). Telemetry-derived 7-lead and single-lead ECGs underestimated the corrected QT interval by 13.45±32.05 msec and 19.62±33.19 msec compared to standard 12-lead ECGs (p<0.05).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: