Key result
Neck-worn acoustic sensor algorithm tracks sleep heart rate with ~91% accuracy vs. commercial devices.
Why the study?
Does a novel algorithm using a neck-placed wearable acoustic sensor accurately extract heart rate compared to commercial devices in sleeping subjects?
Observational (n=10)
No
Does a novel algorithm using a neck-placed wearable acoustic sensor accurately extract heart rate compared to commercial devices in sleeping subjects?
A novel algorithm using a neck-placed acoustic sensor originally designed for breathing can accurately extract heart rate, enabling dual vital sign monitoring.
Answers are grounded in this paper and cite their sources.
May enable dual cardiorespiratory monitoring from a single neck sensor in sleep; leaves open validation in larger, diverse cohorts before clinical use.
Phonocardiography is a widely used method of listening to the heart sounds and indicating the presence of cardiac abnormalities. Each heart cycle consists of two major sounds - S1 and S2 - that can be used to determine the heart rate. The conventional method of acoustic signal acquisition involves placing the sound sensor at the chest where this sound is most audible. Presented is a novel algorithm for the detection of S1 and S2 heart sounds and the use of them to extract the heart rate from signals acquired by a small sensor placed at the neck. This algorithm achieves an accuracy of 90.73 and 90.69%, with respect to heart rate value provided by two commercial devices, evaluated on more than 38 h of data acquired from ten different subjects during sleep in a pilot clinical study. This is the largest dataset for acoustic heart sound classification and heart rate extraction in the literature to date. The algorithm in this study used signals from a sensor designed to monitor breathing. This shows that the same sensor and signal can be used to monitor both breathing and heart rate, making it highly useful for long-term wearable vital signs monitoring.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2015) reported an observational. Wearable acoustic sensor placed at the suprasternal notch with a novel heart rate extraction algorithm vs. Commercial reference devices (SomnoMedics and Konica-Minolta) was evaluated on Percentage of heart rate values within a ±10% error margin compared to reference devices. A novel algorithm extracting heart rate from an acoustic sensor placed at the neck achieved an overall accuracy of 90.73% and 90.69% compared to two commercial reference devices in 10 sleeping subjects.