Answers are grounded in this paper and cite their sources.
Urinary 11-dehydro-thromboxane B2 (11-DH-TXB2) is a diagnostic marker for evaluating the effect of aspirin treatment in patients with cardiovascular disease. Therefore, it is of great clinical significance to develop a rapid detection method for 11-DH-TXB2 in urine. In this study, we prepared an ultrasensitive monoclonal antibody (mAb) against 11-DH-TXB2. The mAb belongs to the IgG1 subtype with kappa light chains. In an indirect competitive ELISA , it had a half-maximal inhibitory concentration of 3.63 ng/mL and a high affinity of 5.60 × 10 9 L/mol for 11-DH-TXB2. We then developed an immunochromatographic strip for point-of-care testing (POCT) of 11-DH-TXB2 in human urine. The strip provides a cut-off value and limit of detection of 500 ng/mL and 31.2 ng/mL, respectively. The result is obtained within 15 min. In summary, the proposed strip is a portable POCT tool for monitoring 11-DH-TXB2 in human urine.
No takes yet. Share an insight, caveat, or question.
Tan et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: