Biventricular pacing ICDs can lead to inappropriate shocks due to double detection of ventricular premature beats, highlighting an inherent detection problem of dual ventricular sensing.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to double-sensing risk in BiV ICDs; leaves open whether algorithms need redesign.
SCHREIECK, J., et al. : Inappropriate Shock Delivery Due to Ventricular Double Detection with a Biventricular Pacing Implantable Cardioverter Defibrillator. This report describes a patient with advanced heart failure, pronounced intraventricular conduction delay, and ventricular tachycardias who underwent implantation of a multisite pacing ICD. Pacing leads were placed in the right atrium, right ventricular apex, and to the left ventricular posterior wall via a coronary sinus vein. The system proved to have correct sensing and pacing function in an atrial synchronized biventricular pacing mode and an appropriate detection of ventricular fibrillation. However, 1 month after implantation the patient received an inappropriate shock delivery due to double detection of ventricular premature beats. The inherent detection problem of dual ventricular sensing is discussed.