Key result
The Corsano 287 bracelet with PPG technology detected heartbeats within 100 ms of ECG with 94.6% accuracy (95% CI 94.1-95.1) in cardiac outpatients.
Why the study?
Wearable photoplethysmography sensors potentially enable accessible rhythm monitoring, but the heartbeat detection accuracy of a state-of-the-art bracelet and its signal qualifier in cardiac patients needed investigation.
Does photoplethysmography (PPG) using a wearable bracelet accurately detect heartbeats compared to resting ECG in cardiac outpatients?
Observational (n=180)
Does photoplethysmography (PPG) using a wearable bracelet accurately detect heartbeats compared to resting ECG in cardiac outpatients?
The Corsano 287 PPG bracelet demonstrates high accuracy for heartbeat detection and HR/RR-interval measurement in cardiac outpatients, particularly when utilizing a signal quality indicator.
High-accuracy PPG HR/RR detection supports wearable monitoring feasibility in cardiac outpatients; leaves open prospective arrhythmia validation.
Introduction Photoplethysmography (PPG) in wearable sensors potentially plays an important role in accessible heart rhythm monitoring. We investigated the accuracy of a state-of-the-art bracelet (Corsano 287) for heartbeat detection in cardiac patients and evaluated the efficacy of a signal qualifier in identifying medically useful signals. Methods Patients from an outpatient cardiology clinic underwent a simultaneous resting ECG and PPG recording, which we compared to determine accuracy of the PPG sensor for detecting heartbeats within 100 and 50 ms of the ECG-detected heart beats and correlation and Limits of Agreement for heartrate (HR) and RR-intervals. We defined subgroups for skin type, hair density, age, BMI and gender and applied a previously described signal qualifier. Results In 180 patients 7,914 ECG-, and 7,880 (99%) PPG-heartbeats were recorded. The PPG-accuracy within 100 ms was 94.6% (95%CI 94.1-95.1) and 89.2% (95%CI 88.5-89.9) within 50 ms. Correlation was high for HR (R = 0.991 (95%CI 0.988-0.993), n = 180) and RR-intervals (R = 0.891 (95%CI 0.886-0.895), n = 7880). The 95% Limits of Agreement (LoA) were -3.89 to 3.77 (mean bias 0.06) beats per minute for HR and -173 to 171 (mean bias -1) for RR-intervals. Results were comparable across all subgroups. The signal qualifier led to a higher accuracy in a 100 ms range (98.2% (95%CI 97.9-98.5)) (n=143). Conclusion We showed that the Corsano 287 Bracelet with PPG-technology can determine HR and RR-intervals with high accuracy in cardiovascular at-risk patient population among different subgroups, especially with a signal quality indicator.
No takes yet. Share an insight, caveat, or question.
Blok et al. (2024) conducted an observational in cardiac patients (n=180). Corsano 287 bracelet vs. resting ECG was evaluated on accuracy for detecting heartbeats within 100 ms of ECG-detected heart beats (95% CI 94.1-95.1). The Corsano 287 bracelet with PPG technology detected heartbeats within 100 ms of ECG with 94.6% accuracy (95% CI 94.1-95.1) in cardiac outpatients.