Key result
Self-monitoring of blood pressure by pregnant women yielded a mean systolic difference of 0.79 mmHg compared to hospital staff measurements (95% CI -0.008 to 1.58; P=0.052).
Why the study?
Does self-monitoring of blood pressure using a semiautomatic device agree with hospital staff measurements in pregnant women?
Cross-Sectional (n=100)
Does self-monitoring of blood pressure using a semiautomatic device agree with hospital staff measurements in pregnant women?
Self-monitoring of blood pressure using a semiautomatic device in pregnant women is feasible and shows acceptable agreement with hospital staff measurements.
Supports self-monitoring feasibility in pregnancy; leaves open larger trials on maternal-fetal outcomes.
OBJECTIVE: The aim of this study was to assess the feasibility of self-monitoring of blood pressure with a semiautomatic device in pregnant women. PARTICIPANTS AND METHODS: Women attending routine obstetrical ultrasound scanning were invited to participate. The hospital staff initially demonstrated and instructed each participant in correct measurement and then took three measurements on both arms. The participant then repeated the measurements and filled an evaluation questionnaire. We used a validated semiautomatic device for all measurements. Mean values were calculated for systolic, diastolic and mean arterial blood pressure (MAP) and were compared using the paired sample t-test. Mean values and differences of systolic and diastolic pressure were plotted in Bland-Altman plots to test the agreement of the measurements. Finally, a mean evaluation score was calculated. RESULTS: One hundred pregnant women were included in the study. Mean values of systolic, diastolic and MAP were 110.6, 69.7 and 83.3 mmHg, respectively, as assessed by the hospital staff. The corresponding self-measurements were 111.4, 70.2 and 83.9 mmHg, respectively. Mean differences between hospital and self-measurements were 0.79 mmHg for systolic [P=0.052, 95% confidence interval (CI)=-0.008 to 1.58], 0.49 mmHg for diastolic (P=0.056, 95% CI=-0.01 to 0.99) and 0.59 mmHg for MAP (P=0.019, 95% CI=0.099-1.08). The mean evaluation score was 9.2 of 10. CONCLUSION: Differences between hospital staff and self-measurements in systolic, diastolic and MAP are within acceptable international standards. The semiautomatic device Microlife-VSA is well suited for self-measurement; however, safety studies on the use of home measurements in hypertensive pregnancies are still warranted.
No takes yet. Share an insight, caveat, or question.
Lihme et al. (2017) conducted a cross-sectional in Pregnancy (n=100). Self-monitoring of blood pressure vs. Hospital staff measurement was evaluated on Mean difference in systolic blood pressure (MD 0.79 mmHg, 95% CI -0.008 to 1.58, p=0.052). Self-monitoring of blood pressure by pregnant women yielded a mean systolic difference of 0.79 mmHg compared to hospital staff measurements (95% CI -0.008 to 1.58; P=0.052).