Key result
Self-initiated automated blood pressure measurement yielded identical average blood pressures (125/77 mmHg) to 24-hour ABPM but had wide limits of agreement (-20 to +23 mmHg for systolic).
Why the study?
Does self-initiated automated blood pressure measurement provide an accurate estimate of ambulatory blood pressure compared to 24-hour ABPM in pregnant women suspected of having white coat hypertension?
Observational (n=66)
No
Does self-initiated automated blood pressure measurement provide an accurate estimate of ambulatory blood pressure compared to 24-hour ABPM in pregnant women suspected of having white coat hypertension?
Self-initiated automated blood pressure measurement cannot reliably replace 24-hour ambulatory blood pressure monitoring for individual clinical management in pregnant women due to wide limits of agreement.
Answers are grounded in this paper and cite their sources.
Cannot reliably replace ABPM for individual management of suspected white coat hypertension in pregnancy; leaves open utility for group-level estimates.
OBJECTIVE: To determine whether self-initiated and recorded automated blood pressure measurement can provide an accurate estimate of ambulatory blood pressure in pregnant women suspected of having 'white coat hypertension'. DESIGN: A prospective observational study. SETTING: Women's and Children's Health Unit, St George Hospital, a teaching hospital of the University of New South Wales. POPULATION: Pregnant women being assessed for possible 'white-coat hypertension'. METHODS: Sixty-six pregnant women who were undergoing 24 hour ambulatory blood pressure monitoring (ABPM) in their home or work environment also measured their blood pressure six times during this interval using a self-initiated automated blood pressure recorder (Omron HEM 705CP). Agreement between awake ABPM and Omron recorded blood pressures was tested by Bland-Altman analysis. MAIN OUTCOME MEASURE: Limits of agreement between blood pressures measured by each device. RESULTS: Average blood pressures obtained by the two devices were identical (125/77 mmHg) but limits of agreement were wide, -20 to +23 mmHg for systolic blood pressure and -9 to +15 mmHg for diastolic blood pressure. CONCLUSION: The Omron HEM 705CP is a useful device for measuring group average blood pressures in pregnant women suspected of having white coat hypertension but cannot reliably replace ABPM for clinical management of individual pregnant women.
No takes yet. Share an insight, caveat, or question.
Brown et al. (2003) conducted an observational in White coat hypertension in pregnancy (n=66). Self-initiated automated blood pressure recorder (Omron HEM 705CP) vs. 24-hour ambulatory blood pressure monitoring (ABPM) was evaluated on Limits of agreement between blood pressures measured by each device. Self-initiated automated blood pressure measurement yielded identical average blood pressures (125/77 mmHg) to 24-hour ABPM but had wide limits of agreement (-20 to +23 mmHg for systolic).