Key result
In patients with LBBB, cardiac resynchronization therapy with biventricular pacing ≥97% significantly reduced the risk of heart failure or death compared to ICD (HR 0.32; 95% CI 0.23-0.44; P<0.001).
Why the study?
Does CRT-D with high biventricular pacing percentages reduce the risk of heart failure or death compared to ICD in patients with LBBB?
RCT (n=1,219)
Does CRT-D with high biventricular pacing percentages reduce the risk of heart failure or death compared to ICD in patients with LBBB?
In LBBB patients, CRT-D provides significant clinical benefit over ICD only when biventricular pacing exceeds 90%, with maximal benefit at ≥97%.
Answers are grounded in this paper and cite their sources.
Supports high biventricular pacing in LBBB for CRT-D; leaves open need for randomized confirmation before practice change.
AIMS: Previous studies on biventricular (BIV) pacing and cardiac resynchronization therapy-defibrillator (CRT-D) efficacy have used arbitrarily chosen BIV pacing percentages, and no study has employed implantable cardioverter defibrillator (ICD) patients as a control group. METHODS AND RESULTS: Using Kaplan-Meier plots, we estimated the threshold of BIV pacing percentage needed for CRT-D to be superior to ICD on the end-point of heart failure (HF) or death in 1219 left bundle branch block (LBBB) patients in the MADIT-CRT trial. Patients were censored at the time of crossover. In multivariable Cox analyses, no difference was seen in the risk of HF/death between ICD and CRT-D patients with BIV pacing ≤90% [HR = 0.78 (0.47-1.30), P = 0.344], and with increasing BIV pacing the risk of HF/death was decreased [CRT-D BIV 91-96% vs. ICD: HR = 0.63 (0.42-0.94), P = 0.024 and CRT-D BIV ≥97% vs. ICD: HR = 0.32 (0.23-0.44), P < 0.001]. The risk of death alone was reduced by 52% in CRT-D patients with BIV ≥97% (HR = 0.48, P < 0.016), when compared with ICD patients. Within the CRT-D group, for every 1 percentage point increase in BIV pacing, the risk of HF/death and death alone significantly decreased by 6 and 10%, respectively. Increasing BIV pacing percentage was associated with significant reductions in left ventricular volume. CONCLUSION: In patients with LBBB, who were in sinus rhythm at enrolment, BIV pacing exceeding 90% was associated with a benefit of CRT-D in HF/death when compared with ICD patients. Furthermore, BIV pacing ≥97% was associated with an even further reduction in HF/death, a significant 52% reduction in death alone, and increased reverse remodelling. Clinical trials.gov identifier: NCT00180271.
No takes yet. Share an insight, caveat, or question.
Ruwald et al. (2014) conducted an RCT in Left bundle branch block (LBBB) (n=1,219). Cardiac resynchronization therapy-defibrillator (CRT-D) with biventricular pacing ≥97% vs. Implantable cardioverter defibrillator (ICD) was evaluated on Heart failure (HF) or death (HR 0.32, 95% CI 0.23-0.44, p=<0.001). In patients with LBBB, cardiac resynchronization therapy with biventricular pacing ≥97% significantly reduced the risk of heart failure or death compared to ICD (HR 0.32; 95% CI 0.23-0.44; P<0.001).
Synapse has enriched 10 related papers. Consider them for comparative context: