Why the study?
Portable wellness devices can aid vital sign tracking, prompting evaluation of the performance of the wireless, compact, noninvasive VitalWellness device.
Does the VitalWellness device accurately measure vital signs compared to FDA-approved reference devices in adult volunteers?
Does the VitalWellness device accurately measure vital signs compared to FDA-approved reference devices in adult volunteers?
Answers are grounded in this paper and cite their sources.
The VitalWellness device shows moderate to high correlation for heart rate and blood pressure but requires improved accuracy for medical applications.
Background and Objective: Portable and readily accessible wellness devices can aid vital sign measuring for those interested in tracking their health. In this diagnostic accuracy study, we evaluated the performance of the VitalWellness device (VW), a wireless, compact, noninvasive device that measures four vital signs (VS) - blood pressure (BP), heart rate (HR), respiratory rate (RR), and body temperature (BT) - using the index finger and forehead. Methods: Adult volunteers with VS that fell both within and outside of the normal physiological range were enrolled to provide BP, HR, RR, and BT measurements using both the VW and Food and Drug Administration-approved reference devices. A subgroup of participants underwent an additional test to analyze the VW’s performance on HR and RR outside of normal physiological ranges. Statistical measurements were plotted on scatter and Bland-Altman plots. Sensitivity analyses to evaluate the VW’s performance by gender, skin color, finger size, and auxiliary activities were performed. Results: A total of 263 participants completed the study. On an average, systolic BP measured using the VW was 10 mmHg lower than that of the reference device (correlation coefficient r = 0.7), whereas diastolic BP was 3 mmHg lower ( r = 0.6), and RR was 2 bpm lower ( r = 0.7). VW HR and BT measurements were, on average, 1 bpm and 0.3°F higher than the corresponding reference measurements ( r = 0.9 and r = 0.3), respectively. Conclusion: The VW device is well-suited for home-based, nonmedical monitoring of HR, RR, and BP. Further improvement in measurement accuracy is required to enable applications for medical use.
No takes yet. Share an insight, caveat, or question.
Polanco et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: