Answers are grounded in this paper and cite their sources.
We have developed a method for measuring the electrocardiogram (ECG) continuously during MR examinations. In contrast to ECG acquisition by wires, our new method is to amplify and convert the ECG into an optical signal directly above the patient's heart. The optical signal is transmitted out of the magnet bore by optical fiber. The small and fixed dimensions of the ECG amplifier avoid interactions with the MR system because of shorter electrical structures and smaller enclosed areas. Tests of the proposed device in a 1.5 Tesla MR system show that continuous and reliable ECG monitoring and sequence triggering are possible.
No takes yet. Share an insight, caveat, or question.
Felblinger et al. (1994) studied this question.