Why the study?
Clinical and radiographic factors associated with CIED lead failure within 10 years of implantation remained to be assessed.
Do smaller pocket lead loops and defibrillation leads increase the risk of lead failure in patients with cardiovascular implantable electronic devices?
Do smaller pocket lead loops and defibrillation leads increase the risk of lead failure in patients with cardiovascular implantable electronic devices?
Younger age, defibrillation leads, and smaller pocket lead loops are associated with an increased risk of CIED lead failure, suggesting that avoiding tight loops during implantation could reduce this risk.
Younger age and defibrillation leads may warrant closer surveillance; leaves open need for prospective validation of radiographic predictors.
Objective To assess the clinical and radiographic factors associated with lead failure by comparing subjects with lead failure within 10 years of implantation with an implant‐year‐matched group without lead failure. Methods A case‐control study with 49 subjects who received Cardiac Implantable Electronic Device (CIED) between January 1, 1999 and July 31, 2008 and developed lead failure within 10 years of implantation in a single center. The control group consisted of subjects (n = 54) with normally functioning leads matched one‐to‐one by implant year. Results Among the failure group, the meantime from implantation to device lead failure was 4.70 ± 2.94 years. Older age at implantation was associated with a lower likelihood of lead failure (Odds Ratio (OR) = 0.28 (75 vs 42 years old), 95% CI 0.12‐0.63, P = .002). A larger smallest loop diameter on the chest radiograph was also associated with a lower likelihood of lead failure (OR = 0.51 (31 vs 14 mm), 95% CI 0.27‐0.97, P = .04). CIED type (defibrillator vs pacemaker) and Ottawa scores were not significantly associated with lead failure. Among lead‐specific parameters, defibrillation lead vs pace‐sense lead was associated with lead failure (OR = 3.91, 95% CI 1.95‐7.81, P < .001). Conclusions Younger age, defibrillation leads, and small lead loops are associated with lead failure in CIEDs. Techniques to avoid tight loops in the pocket could potentially reduce the risk of lead failure and bear important implications for the implanting physician.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: